Agilent Technologies 87075C User's And Service Manual page 111

Multiport test set for e506x network analyzers
Hide thumbs Also See for 87075C:
Table of Contents

Advertisement

Object type
Syntax
Description
Variable
Examples
Related objects
Equivalent key
COM Object Reference
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. STATe
SCPI.SENSe(Ch).CORRection.MULTiport.SELF.
STATe
Property
SCPI.SENSe(Ch).CORRection.MULTiport.SELF.STATe = Status
Status = SCPI.SENSe(Ch).CORRection.MULTiport.SELF.STATe
For channel 1 to channel 4, specifies whether to execute the self calibration of the
test set automatically at user defined interval. To specify the interval, use the
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. TIMer on page 103
command. Note that, if SCPI.SENSe(Ch).MULTiport.STATe on page 107 is off,
self calibration is not executed automatically even if it is turned on with this
command.
Status
Description
Auto Self Cal state
Data type
Boolean
Range
True (-1) or False (0)
Preset value
True
For information on the variable (Ch), see Table 7-2, "Variable (Ch)," on page 91.
Dim Status as Boolean
SCPI.SENSe(1).CORRection.MULTiport.SELF.STATe = True
Status = SCPI.SENSe(1).CORRection.MULTiport.SELF.STATe
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. ONCE on page 101
SCPI.SENSe(Ch).CORRection.MULTiport.SELF. TIMer on page 103
SCPI.SENSe(Ch).MULTiport.STATe on page 107
-
-
[CAL]
Test Set Cal
102
Self Cal
Chapter 7