Advertisement

Anyone who may use, maintain or repair the instrument should read this operating manual carefully and only follow the
operating manual to achieve the design level. Otherwise the instrument will not work properly and will cause malfunction and
damage to the instrument.
Warning:
The readings beyond the range indicate the concentration of the gas that may or will soon reach the explosion concentration or
the concentration of casualties
Replacing components can affect the intrinsic safety of the instrument.
It is strictly forbidden to charge the meter in a hazardous area. Users must use the special charger equipped with this unit.
Do not disassemble, recharge or replace batteries in hazardous locations.
Charging must be turned off; the instrument must not be operated while charging or exporting data.
Portable products, before entering the danger zone, doing electrostatic release of human body, and then carries the instrument
into the zone.
Do not use the computer's USB interface for communication in hazardous areas.
Old and new batteries or batteries from different manufacturers cannot be mixed.
Attentions:
To ensure maximum safety and performance, please read and follow the items and conditions listed below.
 The air intake of the sensor must be kept clean. Blocking the sensor's air intake may result in readings that are lower than
the actual gas concentration or even the gas concentration. It is strictly forbidden to use compressed air to clean the air
inlet of the sensor to avoid damage to the sensor.
 Catalytic Principle Flammable Gas Sensor and Infrared Principle Sensor can only accurately monitor designated gases. It
should be noted, however, that the catalytic principle flammable sensor is not specific to a particular gas and will be able to react
to other flammable gases.
 If the instrument is subject to physical impact or high pollution, the calibration cycle needs to be shortened.
 The sensitivity of flammable gas sensors will be affected by high concentrations of sulfides, halogen compounds, silicon-
containing compounds, and gases containing "mercury" and "lead", steam, and high temperature organic acid gases. Avoid
using them in environments containing the above chemicals. The instrument, if necessary, must be tested and calibrated after
use to avoid affecting the accuracy of the instrument.
 The instrument is strictly prohibited from being used in a high concentration of flammable gas for a long time to avoid damage
to the flammable sensor. If it is necessary, the instrument should be tested and calibrated after use to avoid affecting the
accuracy of the instrument. If the flammable sensor is damaged, it needs to be replaced with a new one. The sensor can only
be reused after it has been calibrated.
 Do not use organic solvents, soap or silicon-containing solutions to clean the instrument to avoid damaging the sensor.
 When sampling with a sampling tube, do not use a sampling tube made of silicone material or other sampling tubes that are
not certified by the special source.
 A sudden change in air pressure or a blocked air circuit may cause temporary fluctuations in the instrument's oxygen readings.
 Large changes in pressure will result in incorrect readings. If the measured gas pressure is greater than 10% of the atmospheric
pressure, it will affect the instrument reading and even damage the sensor. The instrument should be re-calibrated and the
measured gas should be depressurized.
www.sisco.com
GD-600 User Manual
Please read carefully before using
1
sales@sisco.com
+86 773-280-9002

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SKY8000 and is the answer not in the manual?

Questions and answers

Summary of Contents for SISCO SKY8000

  • Page 1: Sky8000 Gas Analyzer User Manual

     Large changes in pressure will result in incorrect readings. If the measured gas pressure is greater than 10% of the atmospheric pressure, it will affect the instrument reading and even damage the sensor. The instrument should be re-calibrated and the measured gas should be depressurized. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 2  This instrument has the function of recording and querying the calibration date. Please check it frequently. Not only do users need to understand the parameters of the instrument, but they also need to understand the meaning of the test data obtained. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 3: Table Of Contents

    Contents SKY8000 Gas Analyzer User Manual ......................... 1 Foreword ..................................1 1. Product Description ..............................4 2. Instrument Characteristics ............................4 3. Attachment configuration list .............................5 4. Technical Parameters .............................. 5 5. Product Description ..............................6 6. Display ..................................7 7. Instructions ................................8 1) Turn on the instrument ............................
  • Page 4: Product Description

    1. Product Description The SKY8000 is a portable gas analysis instrument. It is widely used in industrial waste gas, air pollution source emission detection, boiler flue gas, automobile exhaust gas, indoor air analysis and other applications. The analytical instrument can flexibly detect 1-18 kinds of gases;...
  • Page 5: Attachment Configuration List

    32 hours, 6-in-1 gas can work for more than 24 hours Language Chinese and English bilingual switching Explosion-proof II 2G Ex ia IIC T4 Ga IP66 certification Product Standards Q/WX005-2017 Certificate Measuring instrument license, ATEX explosion-proof certificate, CE certification / EMC certification, ISO certification www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 6: Product Description

    ⑭ Charging interface: Charge the instrument. ⑮ Printer: Print gas concentration information ⑯ Exhaust port: can be connected to the hose, measuring gas discharge interface ⑰ Air inlet: take the sampling tube, measure the gas input interface www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 7: Display

    Min: represents the real-time minimum concentration value since the instrument was turned on. Real-time curve: represents the real-time concentration curve of the current gas of the instrument Relative time: represents the time of the real-time curve relative to the origin www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 8: Instructions

    (Figure 10), two in one (Figure 11), three in one (Figure 12), four in one (Figure 13), five in one (Figure 14), six above the gas (Figure 15) Figure 10 Figure 11 Figure 12 www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 9: Turn Off The Instrument

    (Figure 18), after which the instrument goes to sleep. When the low battery alarm is triggered, the instrument should be turned off and charged as soon as possible, and the charging time is about 4-5 hours. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 10: Gas Detection

    Indicator light + buzzer Disable all bright 50ms,Off 100ms, High Alarm bright 50ms,Off 100ms sound 50ms,Off 100ms sound 50ms,Off 100ms bright 50ms,Off 500ms, Low Alarm bright 50ms,Off 500ms sound 50ms,Off 500ms sound 50ms,Off 500ms Table 1 www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 11: Menu Operation

    Green indicates that the channel is normal, yellow indicates that the channel is low, and red indicates that the channel is high (Figure 21). Figure 21 7) Menu operation The following is a switch map of the menu tree in the instrument (Figure 22) Figure 22 www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 12: Main Menu Interface Operation

    2. When performing this option operation function, please put the instrument in fresh air for 3-5 minutes, until the value displayed by the instrument is stable, otherwise it will cause the display data to be inaccurate. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 13: Concentration Calibration

    2. Setting value: the actual concentration of the current standard gas is input. 3. When selecting the calibration point, the principle of " Zero point < First point < second point < Third point" should be followed, otherwise the instrument calibration will fail. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 14: Alarm Setting

    Iight+ buzzer: When an alarm is triggered in one or more channels, the indicator and buzzer will alarm together in Table 1. All Disabled: When an alarm is triggered in one or more channels, the indicator and buzzer will not generate alarm information. Figure 35 www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 15: 3) Alarm Mode

    Comfort Beep function is turned on. The instrument is in the normal gas detection interface. The instrument is not in the state of battery alarm. The instrument is not in the state of gas alarm. Figure 39 www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 16: Recovery Setting

    Note: 1. When the storage function is turned off, the data cannot be stored. 2. Data storage only supports the storage under the detection interface and real-time curve interface, while other interfaces do not store data. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 17: 1) Data View

    Data Storage interface. Figure 50 In the history data interface, a page supports viewing 10 groups of data, and the cursor can be moved by the direction keys of < and > to select pages to view data. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 18: 3) Historical Curve

    In the menu interface, use the four direction keys of “∧”、“∨”、“<”、“>”to move the cursor to select the unit settings option, and then press the OK key to enter the unit settings sub-menu (as shown in Figure 56); in the sub-menu interface, press the ESC key to return to the menu interface. Figure 56 www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 19: Monitor Settings

    Use the “∧”、“∨” keys to modify the time and seconds parameters of the year, month, day and day. Then press the "OK" key to set the parameters successfully. On the contrary, press "ESC" to cancel the settings. www.sisco.com +86 773-280-9002...
  • Page 20: 2) Background Setting

    10 minutes, 20 minutes, 30 minutes, 45 minutes, 60 minutes, 90 minutes and 120 minutes. The system will automatically enter the automatic shutdown interface (Fig. 61) without any key operation during the set shutdown time, and then perform the shutdown operation. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 21: Others

    Sensor info: information indicating the lifetime, calibration time and recommended replacement time of the sensor can be viewed. Maintenance record: Repair record of current instrument can be viewed. Calibration record: indicating that the last calibration time and the recommended next calibration time can be viewed. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 22: 1) Sensor Information

    "OK" key to enter the "Maintenance Record" sub-menu (as shown in Figure 66); in the sub-menu interface, press the "ESC" key to return to the main menu interface. Figure 66 Maintenance record information of the instrument is displayed in the menu of Maintenance Record. The information in this menu is not modifiable. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 23: 4) Calibration Record

    Insert the tube into the top as far as possible to prevent leakage or external gas flow into the intake, resulting in inaccurate measurement or errors. Figure 70 Figure 71 Tighten the nut clockwise and lock the PU pipe (Fig. 71) www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 24: Common Faults And Solutions

    3. Please check whether the printer port is blocked. 4. Please check that the wiring connection connecting the printer is in good condition. If none of the above problems are solved, please contact the distributor or manufacturer for maintenance. www.sisco.com +86 773-280-9002 sales@sisco.com...
  • Page 25: Terms Of Service

    2. Damage caused by violation of operation rules and requirements. 3.Damage caused by flood, fire and other natural disasters. 4.Damage caused by bad operating environment. 5.The product shall be repaired, modified, modified or disassembled by unauthorized service personnel. www.sisco.com +86 773-280-9002 sales@sisco.com...

This manual is also suitable for:

Gd-600

Table of Contents