8.0 Environmental - Storage and Operating Conditions

The Hoist is intended for internal use within normal environmental conditions.
It is not intended to be used in environments where there are rapid changes in the environmental temperature and
humidity during intended use.
The TX Advanced Ceiling Track Hoist suffers little from any effects of lint, dust and light.
Lint – Due to the nature of the TX Advanced being installed closely to the ceiling, very little lint would be likely to
gain access into the hoists workings. The hoist is recommended as per Service Guide to be wiped cleaned during
every hoist inspection.
Dust – Due to the nature of the TX Advanced being installed closely to the ceiling, very little dust would be likely
to gain access into the hoists workings.
Light – The User controls have been designed to be easily recognisable and the use of bright colours will help the
user through all ranges of lighting. The Specification of the hoist dictates that normal use would occur during
ambient luminance 50 – 500 lux. Additional as the hoist is designed for indoor use only, if required the user may
wish to switch on room lighting. The LCD display on the hoist is backlit to aid with user interaction.

8.1 Normal operating conditions

+5°C to +40°C (41°F to 104°F) at a relative humidity between 15% to 90% RH, non-condensing but not requiring a water vapour
pressure greater than 50hPa and atmospheric pressure between 700hPa to 1060hPa

8.2 Shipping and storage conditions

-25°C to +5°C (-13ºF to 41ºF) with any humidity level.
+5°C to +35°C (41°F to 95°F) at a relative humidity up to 90%.
+35°C to 70°C (95⁰F to 158⁰F) non-condensing at a water vapour pressure up to 50hPa.
12 Hours are required for the Hoist to cool from the maximum storage temperature until ready for its intended use when the
ambient temperature is 20°C (68°F).
12 Hours are required for the Hoist to warm from the minimum storage temperature until ready for its intended use when the
ambient temperature is 20°C (68°F).