Features - AND AD-1690 Instruction Manual

Leak tester
Hide thumbs Also See for AD-1690:
Table of Contents

Advertisement

2. FEATURES

This Leak Tester judges a leak by a pressure change, after the initial pressure is set
inside the instrument, to a maximum of -20 kPa ± 4 kPa.
The Leak Tester can verify a leak in a small instrument easily (Example: Micro pipette.
etc).
The amount, -20 kPa, attained by evacuating the air with the Leak Tester, is approximately
0.2 atmospheres (based on atmospheric pressure at sea level).
(100 kPa is approximately 1 atmosphere (based on atmospheric pressure at sea level).)
This Leak Tester is a tool for judging leakage, and it cannot measure pressure values
accurately. The Leak Tester displays -20 kPa as a reference pressure and measures
leakage as a pressure change from this amount.
This Leak Tester is designed to test the leakage of a pipette or dispenser and prevent
foreign particle from invading the Leak Tester when evacuating the air at the nozzle of
the instrument.
This Leak Tester is protected against invading dust by an air filter located on the main
unit bottom side.
The accessories provide attachments for four different pipettes tip sizes.
You can select an attachment where the tip and pipette size match up.
With the Micro pipette, the available test capacity is up to 10000 μℓ.