AND AD-1690 Instruction Manual page 14

Leak tester
Hide thumbs Also See for AD-1690:
Table of Contents

Advertisement

(3) Changing the monitor time
The leak tester judges the leak by the monitor time.
(Unit: second)
Changeable by 0.5 seconds up to 100 seconds, and
by 10 seconds over 100 seconds up to 990 seconds.
(Factory setting: 3.0 seconds)
Change the value by using the UP or DOWN key.
* The value displayed with the " " mark (left side
of the display) is the value in memory. But while
changing the displayed value, the "
turned off.
Press the SET key to store the value.
After setting finishes, "
for approximately 1 second (the "
(4) End of setting
The Leak Tester returns to the measurement standby
mode by non-operation, or by pressing the SET
key, to memorize the setting value changed with "(2)"
and "(3)" above.
(5) Initializing the setting value
When displaying* "
at once.
* Refer to "(3) Changing the monitor time".
After displaying "
If you want to initialize, press the
UP or DOWN key to change
to " ", and press the SET key.
After the setting value returns to the
value initialized (quantity of
pressure change: 0.3kPa, monitor
time: 3.0 seconds), the display
returns to measurement standby
mode ("
"display).
If you do not want to initialize, when displaying the
"
", press the SET key.
The display returns to measurement standby mode
("
"display) after displaying "
" is displayed
" flashes).
" of (3), press the DOWN key
", "
" is displayed.
Initialize
confirming display
".
Press
" mark is
SET
Not
initialize
14
UP
DOWN
kPa
Store
SET
SET
REDY
SET
Press
DOWN
SET
SET
Press
UP
DOWN
SET
Press
SET
Initialize
S E C