8.3.3. NECESSARY MAINTENANCE

Check accuracy of pressure gauges
Manifold, pressure regulator, valve and gauges: check condition, function and labeling
Check labelling
Check for corrosion
Check function
Pressure test with 1-times working pressure for 12 hours
Worn and defective components must be changed immediately from authorized qualified company
Valve is designed according to ISO 10297, including type test with cycle test from up to 2000 cycles.
Change inner parts after 2000 cycles, change must be done from authorized qualified company
If there are any leakages or too much corrosion at the manifold/ pressure regulator or valve, component
must be replaced with all connections by an authorized company
After the changing of components or tubes, pressure and leakage test must be done again and be proto-
colled

8.4. MEASURES FOLLOWING MAINTENANCE

THE FOLLOWING STEPS MUST BE PERFORMED WHEN THE MAINTENANCE WORKS HAVE BEEN COM-
PLETED AND BEFORE SWITCHING ON THE SYSTEM.
1. Make sure that all tools, materials and other equipment have been removed from work area.
2. Clean the work area and remove any spilled substances, e.g. liquids, processing material or similar.
3. Make sure that safety relevant components are working perfect.

9. TROUBLESHOOTING

The following section describes possible causes of malfunction and how to eliminate them. If malfunctions
occur with increasing regularity, shorten the maintenance intervals to reflect the actual load. If malfunctions
occur that cannot be eliminated with the following help, please contact the manufacturer.

9.1. SAFETY NOTES FOR TROUBLESHOOTING

In any case of malfunction, the shut-off valves oft he connected gas cylinder, gas container or bundle must be
closed. Never operate devices with malfunctions under pressure.
WHAT TO DO IN CASE OF MALFUNCTION:
1. Interrupt and ensure gas supply.
2. Depressurize pressure regulators and close valves.
3. Only trained technical staff may eliminate malfunctions.
4. Restore component to its original state.
Description
Reason
No Flow
Pressure
(Inlet Gauge
Regulator is
shows pres-
closed
sure, outlet
gauge not)
No Flow (Inlet
Cylinder Valve
and Outlet
closed Pro-
Gauge show
cessgas Valve
„0")
closed Pres-
sure Regulator
closed Cylin-
der coil/ hose
not connected
No Flow
Handwheel
closed
Solution
Valve
Open pressure
regulator slowly
Close purge
valve
Initial Start- Up
see point 6
Installation and
initial start-up
Open hand-
wheel slowly
17/44
Cylinder
Line
Gas
Regulator
Regulator
Supply Panel
EN