Summary of Contents for Minebea Intec PR 40 Series
Page 1
Installation Manual Aluminum Single Point Load Cells PR 40/.., PR 43/.., PR 47/.. Translation of the Original Installation Manual 9499 053 04000 Edition 1.6.0 07/21/2023 Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany Phone: +49.40.67960.303 Fax: +49.40.67960.383...
Page 2
Any information in this document is subject to change without notice and does not represent a commitment on the part of Minebea Intec unless legally prescribed. This product should only be operated/installed by trained and qualified personnel. In correspondence concerning this product, the type, name, and release number/serial number as well as all license numbers relating to the product have to be cited.
WARNING indicates that death or severe, irreversible injury may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. CAUTION Warning of personal injury. CAUTION indicates that minor, reversible injury may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Minebea Intec EN-3...
NOTICE indicates that damage to property and/or the environment may occur if appropriate safety measures are not observed. Take the corresponding safety precautions. Note: User tips, useful information, and notes. Hotline Phone: +49.40.67960.444 Fax: +49.40.67960.474 eMail: help@minebea-intec.com EN-4 Minebea Intec...
If there are grounds for rejection of the goods, a claim must be filed with the carrier immediately. The Minebea Intec sales or service organization must also be notified. Before operational startup NOTICE Perform visual inspection.
OIML R76 PR 47/..: 800 mm × 800 mm max. platform size for PR 43/.. max. capacity E Platform size 15 kg, 30 kg 450 mm × 450 mm 50 kg, 100 kg 600 mm × 600 mm EN-6 Minebea Intec...
Warning of hazardous area and/or personal injury When used in a dust ex zone, the risk of electrostatic charging must be minimized. Dust layers on the load cell >5 mm are not permitted. The load cell must be installed securely. Minebea Intec EN-7...
Aluminum Single Point Load Cells PR 40/.., PR 43/.., PR 47/.. 3 Specifications Dimensions PR 40/.. all dimensions in mm PR 43/.. all dimensions in mm EN-8 Minebea Intec...
Page 11
3 Specifications Aluminum Single Point Load Cells PR 40/.., PR 43/.., PR 47/.. PR 47/.. all dimensions in mm Minebea Intec EN-9...
200 kg ..C3MR ..C3MRE ..III5000S PR 47/250 kg 250 kg ..C3MR ..C3MRE ..III5000S PR 47/300 kg 300 kg ..C3MR ..C3MRE ..III5000S PR 47/500 kg 500 kg ..C3MR ..C3MRE ..III5000S PR 47/750 kg 750 kg ..C3MR ..C3MRE ..III5000S EN-10 Minebea Intec...
Output impedance between measuring terminals 350 ±25 Ω Insulation impe- between measuring circuit and housing >5000 MΩ dance at U = 100 V Insulation voltage between circuit and housing Recommended to hold the specified performance ≤12 supply voltage Minebea Intec EN-11...
Page 14
Accuracy classes and min. scale interval of the load cells v for PR 40/20 kg…PR 40/50 kg PR 40 Type Divisions 20 kg 30 kg 40 kg 50 kg Unit OIML C3MR 3000 1.34 2.00 2.67 3.34 NTEP Class III 5000 1.34 2.00 2.67 3.34 5000 Single EN-12 Minebea Intec...
Page 15
Accuracy classes and min. scale interval of the load cells v for PR 47/300 kg…PR 47/750 kg PR 47 Type Divisions 300 kg 500 kg 750 kg Unit OIML C3MR 3000 20.00 33.34 50.00 NTEP Class III 5000 5000 20.00 33.34 50.00 Single Minebea Intec EN-13...
NOTICE Force shunts may cause measuring errors. All incoming and outgoing lines (hoses, pipes, cables) must be coupled to the measured object as flexibly as possible. EN-14 Minebea Intec...
WARNING When installing in potentially explosive atmospheres: It is imperative that you follow the application-dependent installation instructions! Always check whether it is permissible to bilaterally connect the screens to the equipotential bonding. Minebea Intec EN-15...
However, despite the robust nature of the materials used, the cable should be protected from excessive chemical and mechanical stresses. Preventing water from penetrating the end of the cable is also important "life insurance" for the system. EN-16 Minebea Intec...
Some cleaning agents may not be compatible with the load cell material. When using cleaning agents, ensure that their compatibility with the load cell material has been tested and approved (see Chapter 3.2). Clean the load cell with a damp cloth to avoid electrostatic charging. Minebea Intec EN-17...
Minebea Intec at a charge. This option of returning the product is intended to provide proper recycling or reuse in a manner that is collected separately from municipal waste.