Table 5: Error messages
Error-No.:
Description
1
Imbalance arose
2
Imbalance sensor is defective
4
Imbalance switch has been activated for longer than 5 seconds
8
Transponder in the rotor is defective
Leap of speed is too big between two measurements
14
CLOSE lid
33
Open lid while motor is running
40
Communication with frequency converter disturbed during start
41
Communication with frequency converter disturbed during stop
42
Short circuit in the frequency converter
43
Undervoltage frequency converter
44
Overvoltage frequency converter
45
Over temperature frequency converter
46
Over temperature motor
47
Over current frequency converter
48
Timeout between control unit and frequency converter
49
Other error frequency converter
55
Overspeed
99
Rotor is not allowed in this centrifuge
Inserted rotor does not exist in the program
FALSE
Rotor is not detected
rotor no
© Hermle Labortechnik GmbH
manual.docx
Z306 Instruction
APPENDIX