Application Method - Haier HYC-410 Operation Manual

Pharmaceutical refrigerator

Advertisement

Control Panel
 
Compressor
Defosting
Alarm
Silence
Light
Upper
Set
  Test
12

Application method

Temperature display
Caution
The displayed temperatures indicate upper and lower temperatures of the refrigerator perceived
Power
by the sensor in glycerin bottle. It is not always of 5 ℃ . Temperature sensor indicates average
temperature inside the refrigerator.
Alarm
TEMP. SET
Lower
If you want to reset the temperature, for example, 5℃ is set originally and you want to change the
temperature to 5.5℃ , please operate according to the following steps:
TEMP. SET
Key Operation
1
2
Press "Set" for 5s at the same time.
3
Adjust the parameter value by pressing
4
the "Upper" and "Lower" key,pressing
the "Set" key to save.
Defrost Cycle
Forced defrost cycle
When the accumulative startup time of the compressor reaches a certain value (8 hours by
default),the equipment will start the forced defrost cycle. Once the frost is thawed, the refrigerator will
resume normal operation. The temperature in the box may rise during the defrosting cycle.
The temperature in the refrigerator displays.
Click "Set"
The current parameter value fl ashes.
The temperature in the refrigerator displays.