Table 3 – Proximity fields from RF wireless communications equipment
Test frequency (MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
Table 4 - Input a.c. power Port
Phenomenon
Electrical fast transients/burst
Surges Line-to-line
Conducted disturbances induced by RF
fields
Voltage interruptions
Table 5 - Signal input/output parts Port
Phenomenon
Electrostatic Discharge
Rev. B
English
Band (MHz)
380 - 390
430 - 470
704 - 787
800 - 960
1700 - 1990
2400 - 2570
5100 - 5800
Basic EMC standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
Compliance
IEC 61000-4-2
Immunity test levels
Home healthcare environment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
Immunity test levels
Home healthcare environment
±2 kV
100kHz repetition frequency
±0.5 kV, ±1 kV
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º
and 315º
0% UT; 1 cycle and
70% UT; 25/30 cycles Single phase: at 0º
0% UT; 250/300 cycles
UT=rated input voltage
Electromagnetic environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air