  • Page 2 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Error Report The Error Report menu op5on lists all the Errors on the device in Chronological order along with 5me stamp and Error code. This list helps Service personnel to iden5fy and triage an error condi5on on your device.
  • Page 3 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 \\atrion.int\quest-common\Documents\200s--Order Fulfillment\281xxx\SK281007 Rev 1.docx...
  • Page 4: Alarm Overview

    MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Alarm Overview Alarms and errors are the advisories to the operator to ensure proper and safe working of the MPS3 System. The message states the reason for the alarm and contains prompts for clearing the alarm. All Alarms are recorded in the Alarm Report and may be accessed from the Menu op5on.
  • Page 5 flow limit values exceed the set, or the Ischemic 5mer has expired. General Troubleshoo/ng Guide Each MPS 3 System is rigorously tested prior to shipment to ensure the devices func5on reliably as intended. It is expected that over the life of the device, some issues may arise as a result of use condi5ons, environment, handling and storage.
  • Page 6 If using Hypothermic Reservoir: o Ensure the Hypothermic Reservoir has water filled to the drain port. o Elevate the Hypothermic Reservoir above the MPS 3 System so that the outlet port of the reservoir is above the heat exchanger. o Check the water lines to ensure they are properly connected to the correct inlet and outlet ports o Check the water lines to ensure they are not kinked or blocked.
  • Page 7 In order to use the Drain HEX func5onality, the system must be in cold mode and the circula5on pump must be turned off. If a heater-cooler unit is being used with the MPS 3 System, the heater-cooler unit flow must be •...
  • Page 8 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Arrest Cartridge detected in Addi5ve pump (yellow alarm): Make sure the addi5ve cartridge is installed all the way (turned as far as possible). One of the sensors • could poten5ally not be triggered despite the correct cartridge being on the console. This error can falsely occur if the addi5ve cartridge is removed slowly.
  • Page 9 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Error Code List and Solu/ons The following table allows to determine correc5ve ac5on by searching for the screen text or the error code and iden5fying a possible cause and ac5on. The error code can be seen just below the Cau5on or Warning symbol in the message window.
  • Page 10 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Alarm Message Display Correc/ve Ac/on Code Pneuma/c leak or compressor malfunc/on * Connect medical air Consider a\aching medical air if available * Call Service following procedure Consult Service acer this case SHUTDOWN RESTART...
  • Page 11 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Alarm Message Display Correc/ve Ac/on Code * Con5nue with case while manually Air in line sensor test failure monitoring for air in delivery line Air detec/on may be compromised * Call Service following procedure...
  • Page 12 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 *Check the lines for clamps, kinks or occlusions * Manually actuate delivery and vent valves several 5mes. System pressure sensor error * Check heat exchanger to make sure it RETEST is secured 5ghtly with the locking...
  • Page 13 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check the lines for clamps, kinks or occlusions * Manually actuate delivery and vent Excessive system pressure valves several 5mes. Check lines for clamps * Check heat exchanger to make sure it...
  • Page 14 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Manually actuate valve several 5mes. Antegrade delivery valve sensor mismatch * Clean valves if dirty Flow may be resumed CONFIRM * Manually actuate valve several 5mes. Retrograde delivery valve sensor mismatch * Clean valves if dirty...
  • Page 15 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 \\atrion.int\quest-common\Documents\200s--Order Fulfillment\281xxx\SK281007 Rev 1.docx...
  • Page 16 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check delivery line for clamps, kinks, Delivery occlusion occlusions Check for clamps * Check the upper pressure limit Check pressure limit sekng se`ngs CONFIRM * Check blood/crystalloid source lines and pressures Unable to maintain flow rate sekng 91, 92 * Stop flow and restart flow...
  • Page 17 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check arrest delivery line for clamps, kinks, occlusions * Check stopcock posi5on Arrest occlusion detected * If problem persists, replace arrest Check Arrest line and stopcock delivery line and try again Arrest pump is Off...
  • Page 18 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check addi5ve delivery line for clamps, kinks, occlusions * Check stopcock posi5on Addi/ve occlusion detected * If problem persists, replace addi5ve Check Addi/ve line and stopcock delivery line and try again Addi/ve pump is Off...
  • Page 19 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Con5nue procedure without temp Circula/on valve sensor mismatch control with the MPS 3 console Manually monitor delivery temperature * Try connec5ng medical air H2O Circ is OFF. Heaters are disabled. * Restart system...
  • Page 20 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Ensure cold water source has adequate water level and flow rate * Check water lines for clamps, kinks, occlusions * Check water line to ensure they are connected properly and not reversed Low Water flow delayed heater diagnos/cs...
  • Page 21 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Ensure cold water source has adequate water level and flow rate * Check water lines for clamps, kinks, occlusions * Check water line to ensure they are connected properly and not reversed Reverse Water flow delayed heater diagnos/cs...
  • Page 22 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 Pneuma/c leak or internal compressor malfunc/on * Connect to medical air Consider a\aching medical air if available * Call Service following procedure Consult Service acer this case CONFIRM Compressor malfunc/on detected * Connect to medical air...
  • Page 23 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check for clamps, kinks or occlusions * Manually actuate delivery and vent valves several 5mes. Max System pressure * Check heat exchanger to make sure it Check circuit is secured 5ghtly with the locking knob...
  • Page 24 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check Addi5ve Cartridge to verify it is properly installed and fully secured to the pump Addi/ve Cartridge not detected * If the cartridge was secure, try Verify Addi/ve cartridge installa/on removing cartridge and reinser5ng it Addi/ve pump is Off...
  • Page 25 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 System maintenance due * Call Service Please schedule System Maintenance CONFIRM System maintenance past due. * Call Service Please schedule System Maintenance CONFIRM * Restart system System Config mismatch Error 10, 11, * Call service a]er mul5ple...
  • Page 26 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 * Check fans and air inlets for Internal temperature Error detected obstruc5ons Check for obstructed air flow 158, 163 * Restart system Unable to proceed. Call for Service ,165,167 * Call Service a]er mul5ple...
  • Page 27 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 MPS 3 Checklist – Troubleshooting MPS 3 Console Change Out Procedure ☐ Error Reports ☐ Data Transfer ☐ Alarm Overview ☐ General Troubleshooting Guide ☐ Alarms that can be caused by user error: ☐...
  • Page 28 MPS 3 Disposable Change Out Procedure ☐ Case History, Case Reports, Error Report ☐ I hereby acknowledge that I have trained the operator on the above MPS 3 elements. Trainers Name Date I hereby acknowledge that I have attended training on the above Phase Two MPS 3 elements and understand the information outlined.
  • Page 29 MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 (800) 627-0226 - (972) 390-9800 \\atrion.int\quest-common\Documents\200s--Order Fulfillment\281xxx\SK281007 Rev 1.docx...
  • Page 30: Revision History

    MPS 3 Troubleshoo/ng Guide SK281007 Revision 1 REVISION HISTORY Revision Description of Change ECO# Originator Date Release EC20243 5/15/20 \\atrion.int\quest-common\Documents\200s--Order Fulfillment\281xxx\SK281007 Rev 1.docx...