Siemens SITRANS F Modbus RTU RS 485 Operating Instructions Manual page 47

Communication modules
Hide thumbs Also See for SITRANS F Modbus RTU RS 485:
Table of Contents

Advertisement

MODBUS regis‐
MODBUS ad‐
ter
dress
4:02209
2208
4:02211
2210
4:02212
2211
4:02213
2212
4:02214
2213
Modbus RTU RS 485
Operating Instructions, 03/2021, A5E00753974-AB
No. of bytes Data type
2
Byte
2
Byte
2
Byte
2
Byte
4
Float
Description
Pulse 1 width:
0 = 64 us
1 = 130 us
2 = 260 us
3 = 510 us
4 = 1.0 ms
5 = 2.0 ms
6 = 4.1 ms
7 = 8.2 ms
8 = 16 ms
9 = 33 ms
10 = 66 ms
11 = 130 ms
12 = 260 ms
13 = 520 ms
14 = 1.0 s
15 = 2.1 s
16 = 4.2 s
Frequency 1 selection:
0 = Off (read only)
1 = Massflow
2 = Fraction A
3 = Fraction B
4 = Volumeflow
5 = Sensor temp.
6 = Density
7 = % fraction A
Frequency 1 Direction
1 = Unidirectional
2 = Bidirectional
Frequency 1 Fmax.
0 = 10 kHz
1 = 5 kHz
2 = 1 kHz
3 = 500 Hz
Frequency 1 Timeconstants
Modbus holding registers
A.6 Output
Read / write
R/W
R/W
R/W
R/W
R/W
47

Advertisement

Table of Contents
loading

Table of Contents