Eeprom Update; Reverification; Accuracy Under Calibration Conditions - National Instruments NI-9269 Calibration Procedure

Hide thumbs Also See for NI-9269:
Table of Contents

Advertisement

EEPROM Update

When an adjustment procedure is completed, the NI 9269 internal calibration memory
(EEPROM) is immediately updated.
If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments by initializing an external calibration,
setting the C Series calibration temperature, and closing the external calibration.

Reverification

Repeat the
Verification
Note
If any test fails Reverification after performing an adjustment, verify that you
have met the
Go for Support

Accuracy Under Calibration Conditions

The values in the following table are based on calibrated scaling coefficients, which are stored
in the onboard EEPROM.
The following accuracy table is valid for calibration under the following conditions:
Ambient temperature 23 °C ± 5 °C
NI 9269 installed in slot 8 of an NI cDAQ-9178 chassis
Slots 1 through 7 of the NI cDAQ-9178 chassis are empty
Table 6.
Gain Error
*
Range equals 10.47 V
Note
For operational specifications, refer to the most recent NI 9269 Operating
Instructions and Specifications online at
6 | ni.com | NI 9269 Calibration Procedure
section to determine the As-Left status of the device.
Test Conditions
for assistance in returning the device to NI.
NI 9269 Accuracy Under Calibration Conditions
0.060%
before returning your device to NI. Refer to
ni.com/manuals
Offset Error
0.015%
.
Where to

Advertisement

Table of Contents
loading

Table of Contents