National Instruments NI 9262 Calibration Procedure
National Instruments NI 9262 Calibration Procedure

National Instruments NI 9262 Calibration Procedure

1 ms/s/ch simultaneous, 10 v, 6-channel c series voltage output module
Hide thumbs Also See for NI 9262:

Advertisement

Quick Links

CALIBRATION PROCEDURE
NI 9262
1 MS/s/ch Simultaneous, ±10 V, 6-Channel C Series Voltage
Output Module
This document contains the verification and adjustment procedures for the NI 9262. For more
information on calibration solutions, visit
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4
Accuracy Verification....................................................................................................... 4
Adjustment................................................................................................................................ 5
Accuracy Adjustment ....................................................................................................... 5
EEPROM Update ..................................................................................................................... 6
Reverification ........................................................................................................................... 6
Accuracy Under Calibration Conditions .................................................................................. 6
NI Services ............................................................................................................................... 7

Software

Calibrating the NI 9262 requires the installation of NI-DAQmx 19.5 or later on the calibration
system. You can download NI-DAQmx from
LabVIEW, LabWindows
to install support for only the application software you intend to use.
ni.com/calibration
/CVI
, ANSI C, and .NET. When you install NI-DAQmx, you need
ni.com/downloads
.
. NI-DAQmx supports

Advertisement

Table of Contents
loading

Summary of Contents for National Instruments NI 9262

  • Page 1: Table Of Contents

    Reverification ........................... 6 Accuracy Under Calibration Conditions .................. 6 NI Services ..........................7 Software Calibrating the NI 9262 requires the installation of NI-DAQmx 19.5 or later on the calibration system. You can download NI-DAQmx from . NI-DAQmx supports ni.com/downloads ™...
  • Page 2: Documentation

    Documentation Consult the following documents for information about the NI 9262, NI-DAQmx, and your application software. All documents are available on , and help files install with the ni.com software. NI cDAQ-9174/9178 USB Chassis Quick Start NI-DAQmx installation and hardware setup...
  • Page 3: Test Equipment

    Test Equipment NI recommends that you use the equipment in Table 1 for calibrating the NI 9262. If the recommended equipment is not available, select a substitute using the requirements listed in Table 1. Table 1. Recommended Equipment Recommended Equipment...
  • Page 4: Verification

    AOx+ AOx– NI 9262 Set the DMM to read DC voltage in the minimum range required to cover the NI 9262 output range of ±10 V. Acquire a sample Create and configure an AO voltage channel according to Table 2.
  • Page 5: Adjustment

    Call the NI 9262 get C Series adjustment points function to obtain an array of recommended calibration voltages for the NI 9262. Set the DMM to read DC voltage in the minimum range required to cover the NI 9262 output range of ±10 V.
  • Page 6: Eeprom Update

    Repeat step 1 for each channel on the NI 9262. EEPROM Update When an adjustment procedure is completed, the NI 9262 internal calibration memory (EEPROM) is immediately updated. If you do not want to perform an adjustment, you can update the calibration date and onboard calibration temperature without making any adjustments by initializing an external calibration, setting the C Series calibration temperature, and closing the external calibration.
  • Page 7: Ni Services

    NI product. Product registration facilitates technical ni.com/register support and ensures that you receive important information updates from NI. NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504, USA. NI 9262 Calibration Procedure | © National Instruments Corporation | 7...
  • Page 8 For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and...