Thermo Scientific 996 Operating And Maintenance Manual page 39

-86c ultra low temperature freezer
Table of Contents

Advertisement

Section 6
Specifications
6-2
900 Series -86C ULT Freezer
Model
906
Temperature Range
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
40.8"W x 77.8" H x 37.0" F-B
Exterior Dimensions
(103.6cm x 197.6cm x 94.0cm)
30.6"W x 51.5"H x 27.0" F-B
Interior Dimensions
(77.7cm x 130.8cm x 64.3cm)
Capacity
23.0 cu. ft. (651.3 liters)
Refrigeration
Two 1 HP (2545 BTUH each)
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 4.5" (11.4 cm)
Insulation
door
230VAC, 50/60 Hz, 12.0 FLA
Electrical
Operating Range: 208VAC-240VAC
15 Amp, 230VAC, Dedicated Circuit,
Breaker Requirements
15 Amp Time Delay Breaker
Shipping Weight
900 lbs. (408.2 kg)
Model
907
Temperature Range
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
46.8"W x 77.8" H x 37.0" F-B
Exterior Dimensions
(118.9cm x 197.6cm x 94.0cm)
36.6"W x 51.5"H x 27.0" F-B
Interior Dimensions
(93.0cm x 130.8cm x 68.6cm)
Capacity
28.0 cu. ft. (792,8 liters)
Refrigeration
Two 1 HP (2545 BTUH each)
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 4.5" (11.4 cm)
Insulation
door
230VAC, 50/60 Hz, 12.0 FLA
Electrical
Operating Range: 208VAC-240VAC
15 Amp, 230VAC, Dedicated Circuit,
Breaker Requirements
15 Amp Time Delay Breaker
Shipping Weight
980 lbs. (444.5 kg)
956
120VAC, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
20 Amp, 120VAC, Dedicated Circuit,
20 Amp Time Delay Breaker