Page 1
INSTRUCTION MANUAL DIGITAL EAR THERMOMETER CTD505 Model: (Antibacterial Model) English Español 1801...
Page 3
Accessory: One +3V CR2032 Lithium Battery Do not immerse CTD505 in water, or any other liquid. – Eng 1 –...
Page 4
The patient is an intended operator. CTD505 is an adjusted mode (ear mode) clinical thermometer. 1. Follow steps in “Cleaning and Disinfection” section to clean CTD505 up. 3. If the CTD505 has saved past results, the LCD screen automatically displays the latest result. If not, skip to next step.
Page 5
Once deleted, your data can not be restored. Small children may choke on battery, or tiny component parts by accident. Note: Please dispose of worn-out device and batteries per local regulations. – Eng 3 –...
Page 6
Afterwards, use the same alcohol swab to clean exterior area back and forth for another 15 seconds (around 20 times). Be sure to clean CTD505 up before disinfection. Please wash your hands with soap and water before disinfection. Signs of Deterioration During Cleaning and Disinfection, beware of following signs of deterioration: If any signs listed above occur, please contact your local distributor.
Page 7
Body Site Normal Range 35.5 ~ 37.5 °C Oral 95.9 ~ 99.5 °F 34.7 ~ 37.3 °C Axillary 94.5 ~ 99.1 °F 36.8 ~ 38.0 °C Rectal 98.2 ~ 100.4 °F 35.8 ~ 37.7 °C 96.4 ~ 99.9 °F Chamberlain, J.M.,Terndrup,T.E., The ears have it. Using and interpreting tympanic thermometers., Advance PA, February 1995.
Page 9
Safety standard: EN 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance EMC standard: EN 60601-1-2 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests Performance standards: EN 12470-5 Clinical thermometers - Part 5: Performance of infra-red ear thermometers (with maximum device)
TYPE BF APPLIED PART Authorized Representative in the European Community: Prinsessegracht 20, 2514 AP The Hague, The Netherlands Date of Manufacture Factory: Health & Life (Suzhou) Co., Ltd. Serial Number No.1428 Xiang Jiang Road, Suzhou New District, Suzhou City 215129, Jiangsu Province, China Batch code –...
Page 11
Maximum 35.5 °C ~ 42.0 °C (95.9 °F ~ 107.6 °F):± 0.2 °C / ± 0.4 °F Permissible Range Other Range:± 0.3 °C / ± 0.5 °F Display Liquid Crystal Display Temperature: 10.0 °C ~ 40.0 °C (50.0 °F ~ 104.0 °F) Operating ambient Humidity: 15 % ~ 95 % R.H.
Page 12
Guidance and manufacturer’s declaration – electromagnetic emissions The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments: Emissions test Compliance Electromagnetic environment – guidance RF energy is used only to maintain device’s operation. RF emissions Therefore, its RF emissions are so low that it’s not likely Group 1...
Page 13
Guidance and manufacturer’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environments listed below, and should only be used in such environments: Immunity IEC 60601 Compliance Electromagnetic test test level level environment – guidance Conducted RF 3V rms Not Applicable Portable and mobile RF...
Page 14
To prove the accuracy of CTD505, we perform clinical experiments at one hospital by professionals on 3 groups of test subjects: new born to 1-year-old for Group A, 1-year-old to 5-year-old for Group B, and 5-year-old or above for Group C.
Page 15
Accesorio: Una pila de litio CR2032 de +3V...
Page 17
Una vez eliminados, los datos no se pueden restaurar. Los niños pequeños pueden ahogarse con la pila o con componentes pequeños por accidente. NOTA: Deshágase del dispositivo y las pilas utilizados según las normativas locales.
Page 18
Después, utilice un bastoncillo humedecido en alcohol para limpiar el área exterior hacia atrás y hacia delante durante otros 15 segundos (unas 20 veces).
Page 21
Normativa de seguridad: EN 60601-1 Equipos electromédicos, Parte 1: Requisitos generales para la seguridad básica y funcionamiento esencial Normativa EMC: EN 60601-1-2 Equipos electromédicos, Parte 1-2: Requisitos generales para la seguridad básica y características de funcionamiento esencial, Norma colateral: Perturbaciones electromagnéticas; Requisitos y ensayos Normativas de rendimiento: EN 12470-5 Termómetros clínicos.
Page 22
PARTE APLICADA DE TIPO BF Representante autorizado en la Comunidad Prinsessegracht 20, 2514 AP The Hague, The Netherlands Fábrica: Health & Life (Suzhou) Co., Ltd. No.1428 Xiang Jiang Road, Suzhou New District, Suzhou City 215129, Jiangsu Province, China Código de lote...
Page 23
35,5 °C ~ 42,0 °C (95,9 °F ~ 107,6 °F): ± 0,2 °C / ± 0,4 °F Otro intervalo: ± 0,3 °C / ± 0,5 °F Pantalla Pantalla de cristal líquido Condiciones Temperatura: 10.0 °C ~ 40.0 °C (50.0 °F ~ 104.0 °F) Humedad: 15 % ~ 95 % R.H.
Page 24
6-1-12, Tanashi-cho, Nishi-Tokyo-shi, Tokyo 188-8511, Japan E-mail: sales-oe@systems.citizen.co.jp http://www.citizen-systems.co.jp/...