Preventive Maintenance - Thermo Scientific 5900 Series User Manual

-86c ult upright freezer
Table of Contents

Advertisement

Section 4
Maintenance
Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your
unit functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum
performance and efficiency, it is recommended that the unit be checked and calibrated periodically by a qualified service technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the instruction manual for
further details.
We have qualified service technicians, using NIST traceable instruments, available in many areas. For more information on Preventive
Maintenance or Extended Warranties, please contact the Technical Services Department. Cleaning and calibration adjustment intervals
are dependent upon use, environmental conditions and accuracy required.
Tips:
Fill an upright by starting at the bottom near the probe and add racks to one shelf at a time. Allow freezer to recover to set point
between shelves.
Fill a chest by starting at the left side near the probe. Filling with room temperature racks will result in a long pull-down time.
Fill unit with frozen product to help overall performance; frozen water jugs, for example.
Always make certain the vacuum relief port is free of frost and ice, to allow for timely re-entry into the freezer after a door
opening.
Action
Verify ambient temperature, <90°F
* Adjust door handle for firm latching, as needed
Check and clean probe cover, gaskets, hinges and lid(s) of ice and snow. See
Figure 1-4 for probe location. See Section 4, Clean Door Gasket.
Check air filter. Clean or replace as needed. See Section 4, Clean Air Filter.
Check alarm back-up battery. See Section 1, Connect Unit to Electrical Power
and Section 4, Replace Battery(s).
Check condenser fan motor for unusual motor noise or vibration.
* Verify and document calibration, at the minimum, annually. See Section 2,
Calibrate.
* Clean condenser compartment and wipe off condenser. See Section 4, Clean
Condenser.
Recommended Maintenance *Back-up System - inject test & battery. See
Section 5.
To minimize ice build-up inside of freezer:
Locate the freezer away from drafts or heating/cooling vents
Keep the number of door openings to a minimum
Minimize the length of time door is open
Make sure door latches securely after opening
4-6
Model 900 and 5900 Series

PREVENTIVE MAINTENANCE

Freezers
Monthly
4
4
4
More frequent cleaning may be
required, depending on use and
environmental conditions.
4
4
Every
Yearly
2 Years
4
4X
**Replace
4
4
4
Thermo Scientific