Advertisement
Please read these instructions for use carefully and observe the information they contain.
Risk to the user.
Note!
Risk to the device.
Important information/tip
Cross-reference with another section.
The temperature measurement varies depending on the part of the body where the measurement is taken. In a healthy person, the variance can be between 0.4°F - 1.8°F (0.2°C - 1°C) in different parts of the body.
Measurements | Thermometer used | |
Forehead temperature | 96.4°F - 99.7°F (35.8°C - 37.6°C) | Forehead thermometer |
Ear temperature | 96.8°F - 100.0°F (36.0°C - 37.8°C) | Ear thermometer |
Oral temperature | 96.8°F - 99.3°F (36.0°C - 37.4°C) | Conventional thermometer |
Rectal temperature | 97.3°F - 100.0°F (36.3°C - 37.8°C) | Conventional thermometer |
Tip
Tip
Taking the body temperature provides a current measurement of a person's temperature. If you are uncertain about interpreting the results or if the values are abnormal (e.g. fever), please consult your doctor. This also applies in the case of slight temperature changes if there are other symptoms of illness such as agitation, severe sweating, flushed skin, fast pulse rate, tendency to collapse, etc.
Press the On/Off button. A beep is output. All segments in the display light up.
The result of the previous measurement is shown in the display for approx. 2 seconds.
Wait until the measured value and the hourglass symbol have disappeared from the display. 2 beeps indicate that the thermometer is ready to take a measurement.
Do not begin measuring until both beeps have been output.
Hold the measuring sensor of the thermometer in the centre of the temple between the eyesocket and the hairline. Correct placement and good skin contact are the prerequisites for reliable measuring values!
The measurement starts automatically as soon as the measuring sensor is placed at the measuring point. A status bar indicates that the measurement is in progress.
To avoid falsifying the measuring result, it is important to remain still during the measurement and not to move the thermometer!
After approx. 6-8 seconds, a long beep is output and the measuring result is shown in the display.
After 1 minute, the device switches off automatically. You can also switch off the thermometer manually by pressing the On/Off button.
The thermometer can display the temperature in both degrees Celsius (°C) and degrees Fahrenheit (°F). To switch between these measurement units, proceed as follows with the device switched off:
Press and hold the On/Off button until the device switches from °C to °F or vice versa.
Display | Cause | Remedy |
![]() | The measured temperature is over 107.6°F (42°C). | Switch the device off and back on again. Repeat the measurement while complying with the instructions for use. If the device again displays "HI", the patient has a high fever. Consult a doctor immediately. |
![]() | The thermometer is too cold or too warm for use. | Ensure that the ambient temperature is within the operating-temperature range of the device, from 61-104°F (16-40°C). Repeat the measurement and observe correct compliance with the measuring conditions. |
![]() | No measured value has been determined because the skin contact was not adequate, movements occurred or the thermometer was not placed correctly. | Please switch off the device and wait 2 minutes. Repeat the measurement and observe correct compliance with the measuring conditions. |
![]() | Malfunction | Do not use the thermometer any longer – contact Customer Service. |
![]() | Battery low | The battery still has enough capacity for approx. 10 measurements – replace the battery soon. |
![]() | Battery empty | It is no longer possible to take any measurements – replace the battery. |
The device requires two AAA (LR03) batteries.
Used batteries should not be disposed of in normal household waste. You are legally required to dispose of the batteries. Dispose of them via your electronics retailer or your local recycling point.
Notes: The codes below are printed on batteries containing harmful substances: Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury. The batteries in this device do not contain any pollutants.
Clean the measuring head and the sensor with an alcohol swab (available from pharmacies) or with a cotton cloth soaked in 70% isopropyl alcohol.
Make sure that no liquid penetrates into the interior of the thermometer.
Clean the housing with a dry cloth. Do not use any abrasive cleaning agents, thinners or benzene, and never immerse the device in water or other cleaning liquids.
After cleaning, please wait 2 minutes before taking a measurement, so that the thermometer can adjust back to room temperature.
The device must not be stored or used at an excessively high or low temperature or humidity (see technical specifications), in sunlight, in association with an electrical current or in dusty locations. Otherwise inaccuracies can occur.
If prolonged storage is planned, you should remove the batteries.
The device must be tested for accuracy (calibrated) after 3 years. For this, please send the complete device to the retailer's or the service address.
Dispose of the device in accordance with EC Directive 2002/96/EC – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities responsible for waste disposal.
If the device is used other than in accordance with the specifications, perfect functioning cannot be guaranteed!
We reserve the right to make technical changes to improve and develop the product.
Measurement method | Contact-measurement technology | ||
Basic function | Body temperature measurement | ||
Measurement unit | Celsius (°C) and Fahrenheit (°F) | ||
Ambient operating conditions | 60.8°F to 104°F (16°C to 40°C) with a relative humidity of 30-85% | ||
Storage conditions | -4°F to 122°F (-20°C to 50°C) with a relative humidity of 15-95% | ||
Measuring area | Temple | ||
Measurement accuracy | 98.6°F - 102.2°F: ± 0.2°F (37°C - 39°C: ± 0.1°C) 96.4°F - 98.6°F: ± 0.3°F (35.8°C - 37°C: ± 0.2°C) <96.4°F: ± 0.5°F (<35.8°C: ± 0.3°C ) 102.2°F - 105.8°F: ± 0.3°F (39°C - 41°C: ± 0.2°C) >105.8°F: ± 0.5°F (>41°C: ± 0.3°C) | ||
Measuring time | Approx. 7 seconds | ||
Memory function | Previous measured value is displayed when the device is switched on. | ||
Acoustic signal | Beep at the end of measurement | ||
Display | Display (16 x 11 mm) | ||
Energy-saving f unction | Device automatically switches off after 1 minute. | ||
Dimensions Width x depth x height | Approx. 12.3 x 3 x 2.4 cm | ||
Weight | 37 g (incl. batteries) | ||
2 x AAA (LR03) batteries | Batteries last for approx. 1,000 measurements | ||
Signs and symbols |
|
Distributed by: Hans Dinslage GmbH, Riedlinger Straße 28, 88524 Uttenweiler, Germany
EU representative: MEDES LTD, 5 Beaumont gate, Shenley hill, Radlett, Herts WD 7, 7AR. United Kingdom
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Sanitas SFT 40, FHT 6 - Forehead Contact Thermometer Manual
Advertisement
Need help?
Do you have a question about the SFT 40 and is the answer not in the manual?
Questions and answers