Firmware Downgrade - Honeywell BW Ultra User Manual

Portable five-gas detector with internal pump
Hide thumbs Also See for BW Ultra:
Table of Contents

Advertisement

Safety Suite Device Configurator (SSDC) - Offline upgrade (use this option if you wish to refresh
the firmware or downgrade the firmware)
This option can be activated in two ways:
1. Having a computer with no internet access or
2. Enabling the Firmware updates using local file option
(Settings -> System Settings -> Local Settings)
Get firmware under RFP format (this format belongs to Safety Suite Device Configurator),
you can get the file in the following site:
and-flame-detection
For the units:
1. Go to the Device Inventory/List and click on the Refresh button to see the connected
devices.
2. There are two ways to select the firmware file:
a. Enable the checkbox next to the device serial number and above a blue toolbar will
appear, select the Update firmware button.
b. Click on the device serial number to open the Device Details.
3. Click the Update button below the Firmware Version section,
4. Select the RFP file to use and click on the Update button.
5. Follow the status of the process on the Notifications queue.
6. Once the message Firmware updated appears on the top, click on the Refresh button again
to see the new firmware version on the devices list.
Note: If there are datalogs stored into the device when you update the FW to Version 1.024
through IntelliDoX and if after 25 minutes the process doesn't finish, open the nest and the
device should show Error 141. Just press the acknowledge button and the device will be
correctly updated.

Firmware Downgrade

Contact Honeywell if a Firmware downgrade is required.
BW Ultra
sps.honeywell.com/us/en/products/safety/gas-
38
User Manual