Invacare® Perfecto₂™Series
SYMPTOM
Alarm:
YELLOW or RED light
Illuminated.
Alarm: Continuous On RED
light
* Only applies to
IRC5PO2AW &
IRC5PO2VAW
GREEN light with YELLOW
Light Flashing.
Alarm:
Rapid
Beep...Beep...
Beep...Beep
30
PROBABLE CAUSE
System Failure:
1.
Low oxygen purity.*
2.
Kinked or blocked tubing, cannula or humidifier.*
3.
Flowmeter set at 0.5 L/min*.
4.
Unit overheating due to blocked air intake.
5.
Insufficient power at outlet.
6.
Internal repairs required.
1.
Internal repairs required.
Potential Obstruction Alert
1.
Possible internal obstruction in the oxygen path.
Kinked or blocked tubing, cannula or humidifier.
2.
Flowmeter set at 0.5 L/min.
SOLUTION
1.
Clean or Replace filters.
2.
Inspect for kinks or blockages. Correct, clean or
replace item. Once corrected, turn power Off for
60 seconds and then turn power back On.
3.
Check flowmeter is set to 1.0 L/min or more.
Refer to Typical Product Parameters.
4.
a.
Remove and clean cabinet filters.
b.
Move concentrator at least 12 in (30,5 cm)
from walls, draperies, and furniture.
5.
DO NOT use extension cords. Move to another
electrical outlet or circuit.
6.
Call Service Provider.
1.
Call Service Provider.
1.
Inspect for kinks or blockages. Correct, clean or
replace item. Once corrected, turn power Off for
60 seconds and then turn power back ON.
2.
Flowrates less than 1 L/min are not recommend.
The use of the pediatric flowmeter will
deactivate the Potential Obstruction Alert.
1195656-A