MMM Group ECOCELL 22 Operating Instructions Manual

Laboratory drying ovens and incubators line standard
Table of Contents

Advertisement

Návod na servis
Operating instructions
Laboratorní sušárny a inkubátory - linie
Standard
Laboratory drying ovens and incubators
line Standard
ECOCELL
ECOCELL
DUROCELL
DUROCELL
VENTICELL
VENTICELL
INCUCELL
INCUCELL V
INCUCELL
INCUCELL V
22, 55, 111, 222, 404, 707
22, 55, 111, 222, 404, 707
22, 55, 111, 222
22, 55, 111, 222
22, 55, 111, 222, 404, 707
22, 55, 111, 222, 222/2, 404,
22, 55, 111, 222, 404, 707
404/2, 707, 707/2
22, 55, 111, 222, 404, 707
22, 55, 111, 222, 404, 707
22, 55, 111, 222, 404, 707
LSI-S_np_en 0911_mmm_V2.05_B2V
LSI-S_ns_cz_0801_V2.01_B2V

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ECOCELL 22 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MMM Group ECOCELL 22

  • Page 1 Návod na servis Operating instructions Laboratorní sušárny a inkubátory - linie Standard Laboratory drying ovens and incubators line Standard ECOCELL 22, 55, 111, 222, 404, 707 ECOCELL DUROCELL 22, 55, 111, 222, 404, 707 22, 55, 111, 222 DUROCELL VENTICELL 22, 55, 111, 222 22, 55, 111, 222, 404, 707 VENTICELL...
  • Page 2 Congratulations on obtaining a new temperature cabinet, designed for uniform tempering of various materials by hot air at a selectable temperature and a selectable time mode in laboratories. There is an air flap installed that enables wet material drying. The temperature course is controlled by an advanced microprocessor (Fuzzy logic) with a digital display and a temperature sensor PT 100.
  • Page 3: Table Of Contents

    CONTENTS: 30. 11. 2009 GENERAL..........................3 1.1 P ..........................3 URPOSE AND USE IMPORTANT INSTRUCTIONS......................3 2.1 U ....................3 NPACKING CHECKING AND TRANSPORT 2.2 B ..........................3 EFORE INSTALLATION 2.3 U ............................4 SEFUL SPACE DESCRIPTION OF THE UNIT......................5 3.1 G ............................5 ENERAL VIEW 3.2 P ....................6 OWER CONNECTION AND CONNECTORS 3.3 C...
  • Page 4 11.5 .........................18 LEXIBLE PT 100 SENSOR 11.6 ..............18 OMMUNICATION SW WARMCOMM FOR PC UNDER WINDOWS 11.7 ..........................18 EPA FILTER 11.7.1 NSTALATION OF HEPA FILTER FILTER DIMENSIONS FUNCTION AND FILTRATION CHARACTERISTIC ........................18 OF THE AIR FILTER 11.8 ...................19 DOOR PASSING THROUGH VERSION 11.8.1 ..................20 ENTICELL 222/2...
  • Page 5: General

    Operating instructions GENERAL IMPORTANT INSTRUCTIONS The electrically heated ovens/incubators are designed for laboratories, especially for tempering of various materials by hot air at UNPACKING CHECKING AND TRANSPORT adjustable temperature and optional time mode. Please check after unpacking whether the An air flap enables to dry wet materials. unit and its accessories are complete and An advanced modern microprocessor (Fuzzy not damaged.
  • Page 6: Useful Space

    Operating instructions and the minimum permitted temperature b) Exhalations from the exhaust outlet must set to 10 °C. be exhausted actively to prevent vapour condensation on the outlet and its dripping The power cord must not come in contact down to the electric part of the device; with hot parts of the units –...
  • Page 7: Description Of The Unit

    12 - mains cord 14 - door sensor X (D) 15 - suction hole (only for VC, ICV, with air flap) 16 - suction hole (with types EC, DC, IC) ECOCELL 22 X (H) X (W) Fig. 1 DESCRIPTION OF THE UNIT...
  • Page 8: Power Connection And Connectors

    Operating instructions 25 Canon connector in the printer Signal The appliances, which are connected with the connector RS-232C, must comply with valid regulations in terms of electric safety and electromagnetic compatibility. We recommend the printer CITIZEN, model iDP 3110-24 RF-A, which is possible to make an additional order of to the apparatus.
  • Page 9: Function And Operating

    Operating instructions program and the indicator light of the initial 15 - indicator light: blinks - set-up mode, active program stage will lit up. burns - the active part of the program is heating-up to present temperature If the temperature in the chamber will be lower 16 - indicator light: blinks - set-up than the preset value, the heating indicator mode, burns - the active part of the...
  • Page 10: Setting Up The Required Values Of Temperature

    Operating instructions f) U6 - Safety system in the function of a off ▼ or ▲, will cause the stop of the actually protective thermostat – maximum permitted running cycle and starting a new cycle with the exposure temperature setting. After pressing pre-selected parameters.
  • Page 11: Protocol Printing

    Operating instructions 2. Set, according tot he chapter User supporting Continuously lighting indicators inform about functions the requested printing interval. 3. the actual active phase of the programme. The heading is printed on the printer; the The permanently lighting indicator lights heading contains the device type, the set provide information on the actually active temperature and selected time interval.
  • Page 12: Safety System Function And Setting

    Operating instructions T exp min – Minimum permitted temperature of SAFETY SYSTEM FUNCTION AND the exposure phase according to the setting in SETTING the user mode U7. The safety system watches the exposure process T delay – maximum allowable temperature of controlled by the control circuit.
  • Page 13: Checking The Proper Function Of The Safety System

    Operating instructions TEMPERATURE UP“ is displayed at the VENTICELL, ECOCELL and DUROCELL beginning of the test, the temperature in the devices cabinet has not dropped to the required one. It If the maximum set permitted temperature in must be let cooled (if the required temperatures the chamber is exceeded, the safety system exceed approx.
  • Page 14: Not Displayed Errors

    Operating instructions setting procedure is correct and corresponds to react. The safety system switched off the safety. the instructions for use. Disconnect the device from the power supply and reconnect relay; the heating elements have Er15 – The upper temperature limit was no current supply.
  • Page 15: Exchanging The Door Sealing And Adjusting The Door

    Operating instructions – sketch position 1. the entire ventilating hole. Hold the shaft of the flap with pliers to prevent turning over. In the apparatus of the volume of 22 there is only one adjustable screw in the side of the door Note: locking device.
  • Page 16: Parameters Of The Unit

    Operating instructions PARAMETERS OF THE UNIT ECOCELL(EC), DUROCELL(DC), VENTICELL(VC),INCUCELL/INCUCELL V (IC/ICV) Technical data volume EC, VC, IC / ICV cca liters Interior of stainless steel material DIN No 1.4301 width depth EC, DC, VC IC / ICV height bearing plate, width x depth Tray 185x265 380x335...
  • Page 17: Electric Connections

    Operating instructions ELECTRIC CONNECTIONS Basic data for connection Mains 230V/50(60)Hz connection 400V/50(60)Hz, 3NPE; 222/2 404/2 707/2 115V/50(60)Hz; 115V/50(60)Hz, 3PE standard types are marked with bold face Mains voltage fluctuation ±10 % Protection against 1410 1410 1410 1410 dangerous contact - class 520x485 520x485 520x485...
  • Page 18: Maintenance

    Operating instructions the chamber. WARRANTY AND In case some contaminated material has escaped into the chamber the user is responsible SERVICE for proper decontamination of all contaminated surfaces with suitable and approved disinfecting Warranty is provided by MMM for satisfactory agent.
  • Page 19: The Way Of Liquidation Of Package And Discarded Unit

    Operating instructions the door. THE WAY OF For types IC, ICV the door with window is not LIQUIDATION OF available. The surface of the door with windows causes PACKAGE AND higher thermal losses, which increases spatial temperature deviations in the chamber DISCARDED UNIT (compared with common door without windows).
  • Page 20: Lockable Door

    Operating instructions 11.3 ● maximal length of the connecting cable is LOCKABLE DOOR 15 m (specified by RS 232 standard) The lock is located on the front door surface. ● free serial port. 11.4 11.7 LEFT DOOR VERSION HEPA FILTER This is a symmetric mirror version, the door can Available only in VENTICELL and INCUCELL be opened to the other side.
  • Page 21: Two Door Passing Through Version

    Operating instructions VENTICELL 22 min 50 Filter class is S (according to DIN 24 184), EU 12 (according to EUROVENT), or H13 (according to EN 1822). HEPA filter installation Insert nuts for HEPA filter attachment in four rectangular holes in the back cover of the device. Put the metal ring and two rubber rings on the suction flue.
  • Page 22: Venticell 222/2 - Installation Data

    Operating instructions b) if you (unlock and) open the door on the on the “loading” side during the exposition, the message “open” is displayed on the display – you have caused the risk of material contamination; after closing the door, the exposition is repeated.
  • Page 23: Double Door Through Version With Automatic Door Blocking

    Operating instructions Detail A min 50 8x (6x) (5x) screw ∅ 4x40 with countersink head with cross groove for fixing to a wall insert dowels with a size of 8 to the wall. Detail B 15x (19x) (15x) screw for a metal sheet ∅ 3,9x9 with a half-round head and a cross groove necessary to screw with cover sheets with holes of ∅...
  • Page 24: Door Blocking Errors

    Operating instructions on the output side and unblocked on the input EMERGENCY DOOR side. Simultaneously with this change, the VENTICELL converts from the End phase into EMERGENCY DOO OPENING the StandBy phase. (The re-blocking could be also made after switching out the device by the (use it only in case of failure or power failure) (use it only in case of failure ON/OFF button in the StandBy phase.)
  • Page 25 Operating instructions 2. To push in the arrow direction and thus to push away the blocking rod (10 -15 mm is enough). In case of standard model – of the right door 2. To push in the arrow direction and thus to push away the blocking rod (10-15 mm is in the arrow direction, in case of the non- standard model –...
  • Page 28 Manufactured in the EU MMM Medcenter Einrichtungen GmbH Semmelweisstrasse 6 D-82152 Planegg Germany T. +49 89 89 92 26 20 F. +49 89 89 92 26 30 e-mail: medcenter@mmmgroup.com...