MMM Group ecocell 22 Instructions For Use Manual

Laboratory drying ovens and incubators
Hide thumbs Also See for ecocell 22:

Advertisement

Instructions for use
Laboratory drying ovens and incubators
line Standard
ECOCELL
DUROCELL
VENTICELL
INCUCELL
INCUCELL V
22, 55, 111, 222, 404, 707
22, 55, 111, 222
22, 55, 111, 222, 222/2, 404,
404/2, 707, 707/2
22, 55, 111, 222, 404, 707
22, 55, 111, 222, 404, 707
LSI-S_np_en_1309_mmm_V2.10_B2V

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ecocell 22 and is the answer not in the manual?

Questions and answers

Summary of Contents for MMM Group ecocell 22

  • Page 1 Instructions for use Laboratory drying ovens and incubators line Standard ECOCELL 22, 55, 111, 222, 404, 707 DUROCELL 22, 55, 111, 222 VENTICELL 22, 55, 111, 222, 222/2, 404, 404/2, 707, 707/2 INCUCELL 22, 55, 111, 222, 404, 707 INCUCELL V 22, 55, 111, 222, 404, 707 LSI-S_np_en_1309_mmm_V2.10_B2V...
  • Page 2 Congratulations on obtaining a new temperature cabinet, designed for uniform tempering of various materials by hot air at a selectable temperature and a selectable time mode in laboratories. There is an air flap installed that enables wet material drying. The temperature course is controlled by an advanced microprocessor (Fuzzy logic) with a digital display and a temperature sensor PT 100.
  • Page 3: Table Of Contents

    CONTENTS: 6. 9. 2013 GENERAL..........................3 ..........................3 urPose and use IMPORTANT INSTRUCTIONS....................3 ....................3 nPacking checking and transPort ........................3 efore installation ..........................5 seful sPace DESCRIPTION OF THE UNIT....................5 ..........................5 eneral view ....................6 ower connection and connectors ..........................7 ontrol Panel FUNCTION AND OPERATING....................7 ......................7 tarting ans switching etting...
  • Page 4 11.6.1 r ..............20 ecording sw warmcomm for Pc under windows 11.6.2 r ..........20 ecording Printing sw Printer archiv for Pc under windows 11.7 ............................20 ePa filter 11.7.1 i nstalation of hePa filter filter dimensions function and filtration characteristic ..........................21 of the air filter 11.8 ....................21 door Passing...
  • Page 5: General

    Instructions for use cultures). The units are designed for temperatures general up to 70 °C optionally up to 100 °C. A quiet operation is characteristic for the version IC The electrically heated ovens/incubators (without a fan), more accurate temperature are designed for laboratories, especially for regulation with small deviations is characteristic tempering of various materials by hot air at for the version ICV (with ventilator).
  • Page 6 Instructions for use With VC and EC types the air exhaust is No articles may be put on he external protected by a cover at the rear wall of the surface of the unit. unit. This cover is placed inside the unit while shipping.
  • Page 7: Seful Space

    Instructions for use difficult. If you open the door in this state, description do not close them until the chamber is cooled down. Otherwise the door of the unit mechanism could be damaged. Regularly – in daily intervals – check by GENERAL VIEW listening if after the start the fan of the case is running.
  • Page 8: Ower Connection And Connectors

    Instructions for use ECOCELL 22 VENTICELL 22 LSI_S_B2v_c_np 1101 V 2.02_TT POWER CONNECTION AND CONNECTORS VENTICELL 22 Fig. 5: Foot of the unit with mains cord - rear view (with the power board I) 1 - Interface for printer 2 - Mains supply...
  • Page 9: Panel

    Instructions for use 25 Canon connector in the printer 15 - indicator light: blinks - set-up mode, burns - the active part of the program is Signal heating-up to present temperature 16 - indicator light: blinks - set-up mode, burns - the active part of the program is time-delayed switching-off 17 - Indicator light: blinks - set-up mode, burns - decrease of temperature...
  • Page 10: Delayed Program Start

    Instructions for use 2. Press the key ON/OFF. The indicator light Select the option of cycles setting by using the above the key, the indicator light of the pre- user menu. 5. By pressing the key X/W change over to the selected program and the indicator light of the initial active program stage will lit up.
  • Page 11: Larm Function

    Instructions for use USER MENU Confirm by ON/OFF pressing. Default setting: Time watching is switched off. Enter the functions by pressing X/W + ▼ h) U9 – Selection of flexible sensor displaying. concurrently. (Yes/No). After you enter the functions, LEDs of all three i) U14 –...
  • Page 12: Nstallation And Setting Of The Printer

    Instructions for use 4.5.1 4.5.2 INSTALLATION AND SETTING OF PRINTING THE PRINTER 1. The printer prints out the heading showing the DPT-6333-V24 does not require any microswitch device type, programme number, number of setting by the user. cycles, pre-set temperature, time of the start delay, and the required time interval.
  • Page 13: Hange Of Maximum And Minimum Values

    Instructions for use D delay - time of the delayed switching may vary independently. If the set maximum temperature and period of time are exceeded, the safety from the set time for ± 2 min. relay switches off the heating, indicates the error D rise - max.
  • Page 14: Ist Of Error Messages

    Instructions for use CAUTION: After you change the value, check by a repeated The safety system function differs according to entry in the U6 mode whether the pre-set value the device type. has been saved. If not, check the correct setting procedure in the instructions for use.
  • Page 15: Djusting And Function Of The Air Flap

    Instructions for use Other error messages The temperature never reaches the required value during the exposure cycle. Disconnect the device from the power supply and then reconnect it. Check whether the mains voltage reaches the If the error repeats, contact the service centre. values required by the Instructions for Use, chapter Mains Connection.
  • Page 16 Instructions for use Návod k použití LSI – linie Standard B2v VENTICELL 22 ExCHANGING THE DOOR VENTICELL 22 SEALING AND ADjUSTING THE DOOR Remove the sealing completely, start in the middle lower part. Fix the new sealing on the edge of the chamber, start in the middle lower part.
  • Page 17 Instructions for use LSI-S_np_en_1309_mmm_V2.10_B2V...
  • Page 18: Parameters Of The Unit

    parameters of the unit ECOCELL (EC), DUROCELL (DC), VENTICELL (VC), INCUCELL/INCUCELL V (IC/ICV) Technical data volume EC, VC, IC / ICV cca liters Interior of stainless steel material DIN No 1.4301 width depth EC, DC, VC IC / ICV height Tray bearing plate, width x depth 185x265...
  • Page 19: Electric Connections

    Instructions for use ELECTRIC CONNECTIONS 222/2+ 404/2+ 707/2+ Basic data for connection : Mains 230V/50(60)Hz connection : 400V/50(60)Hz, 3NPE; 115V/50(60)Hz; 115V/50(60)Hz, 3PE 1410 1410 1410 1410 standard types are marked with bold face ) 520x485 520x485 520x485 920x485 920x485 Mains voltage fluctuation ±10 % Protection against dangerous contact - class :...
  • Page 20: Maintenance

    Instructions for use sequence after it has been cleaned, be careful If the case is placed in a different environment, and slide in the bottom and the sidewalls as far the inspection must be performed in accordance as behind the four projections in the front part of with the local standards.
  • Page 21: The Way Of Liquidation Of Package And Discarded Unit

    Instructions for use you take over the responsibility for possible Caution! damage during the transport and the producer During operation of the cabinet do not touch will reclaim compensation for possible additional the glass surface. This surface cannot be heat- repairs.
  • Page 22: Ockable Door

    Instructions for use 11.2 11.6 ACCESS PORTS OF DIAMETER SUPPORTIVE SW FOR PC 25, 50, 100 11.6.1 RECORDING SW WARMCOMM The access ports are normally located approx. FOR PC UNDER WINDOWS in the middle of the side (right or left) wall of the chamber.
  • Page 23: Function And Filtration Characteristic

    Instructions for use HEPA filter installation Overpressure variant is pre-set in the user mode U13: Overpressure fan is available – “YES”/is not Insert nuts for HEPA filter attachment in four available – “NO”. rectangular holes in the back cover of the device. Put the metal ring and two rubber rings on the suction flue.
  • Page 24: Enticell 222/2

    Instructions for use - as a standard - micro switches of the THE MAIN DIMENSIONS ARE FOR THE TWO-DOOR locking devices of both doors – a display VENTICELL 404; on the panels of both doors reports about THE DIFFERENT DATA IN BRACKETS APPLY TO THE the opening of whichever doors during the TWO-DOOR VENTICELL 222 exposure with the message “open”.
  • Page 25 Instructions for use 11.8.3 – VENTICELL 707/2 INSTALLATION DATA 1) Input power 7200 W 2) Electrical installation – connection of the unit: – plug VDE 0623, DIN 4962/63in, CEE 17, IEC 309, 3P + N + PE, 16A/380-415 VAC – supply to the socket from the switchboard 5x2.5 mm Cu, protection by circuit breaker 16 A –...
  • Page 26: Blocking

    Instructions for use Detail A moment, the door is blocke on the output side 8x (6x) (5x) screw ∅ 4x40 with countersink and unblocked on the input side. Simultaneously head with cross groove for fixing to a wall insert with this change, the VENTICELL converts from dowels with a size of 8 to the wall.
  • Page 27: Emergency Door Opening

    EMERGENCY DOOR OPENING EMERGENCY DO (use it only in case of failure or power failure) (use it only in case of fa EMERGENCY DOOR OPENING Instructions for use 1. To insert a thin object (screwdriver, rod etc.) through the air hole of the door 1.
  • Page 28 2. To push in the arrow direction and thus to push away the blocking rod (10-15 mm is Instructions for use enough). In case of standard model – of the right door in the arrow direction, in case of the non-standard model – of the left door – it would be reversely (always in the direction from handle to the door hinges).
  • Page 29 Date European Directive 2006/95/EC, 2004/108/EC 2012-06-09 Manufacturer MMM Medcenter Einrichtungen GmbH, Semmelweisstrasse 6, D-82152 Planegg / München declare on our own responsibility that product heating oven: electrically heated drying oven intended for heating of laboratory materials in the temperature range up to 300 °C VENTICELL, ECOCELL, DUROCELL, trade name: INCUCELL, INCUCELL V...
  • Page 32 Manufactured in the EU MMM Medcenter Einrichtungen GmbH Semmelweisstrasse 6 D-82152 Planegg Germany T. +49 89 89 92 26 20 F. +49 89 89 92 26 30 e-mail: medcenter@mmmgroup.com...

Table of Contents

Save PDF