4.2 Mounting kit option /R, (order nr.
K1500BY)
This mounting kit replaces the standard
electrode mounting set so that Yokogawa
refillable electrodes with a long glass shaft
can be mounted in a fitting. The mounting
set may be used at process pressures up
to 3 bar. Higher pressure ratings require
the standard mounting set and conse-
quently, electrodes with DIN-dimensions.
Figure 4.2 shows the mounting.
Note:
The pressure on the KCI-solution must
always exceed the liquid pressure.
The refillable electrodes can be connected
to a KCI-reservoir via a silicone tubing
(see figure 4.2). The pressure on the
KCI-solution and consequently, the liquid
outlet of the electrode can be increased by
mounting the reservoir on top of the fitting.
This prevents penetration of the process
liquid into the electrode.
Electrode to be used*
Order nr.
SR20-AC52 Reference electrode
SR20-AS52 Reference electrode
* For specifications see seperate sheets.
Parts and accessories
Order nr.
K1500BY
K1500FZ
K1500GF
K1500GG
IM12B06K02-01E-E_ed09.indd 17
Description
Description
Mounting kit
O-ring for mounting kit
250 ml, KCI-solution (1 m.)
250 ml, KCI-solution (1m.),
thickend
KCI-Reservoir
K1500FU
Silicone tubing
K1500GA
Fig. 4.2 Mounting of refillable electrodes
with a long glass shaft
IM12B06K02-01E-E
17
Protection Hose
K1500CJ/CK
Electrode
O-Rings
K1500BZ