Pasport PS-2194 Instruction Sheet

Pasport PS-2194 Instruction Sheet

Ethanol sensor

Advertisement

Quick Links

Included Parts
1. Amplifier
2. Probe
3. PTFE tape for membrane replacement (not pictured)
Additional Part Required
PASPORT interface or datalogger
Related Parts
Replacement probe (PASCO part 003-10338)
Additional PTFE tape (720-062)
Quick Start
1.
Connect the amplifier to your PASPORT interface.
2.
Connect the probe to the amplifier.
3.
If you are using a computer, connect the PASPORT interface
to it and start DataStudio.
4.
Ensure that the membrane is installed and intact. There
should be no tears or discoloration.
5.
Place the end of the probe in the gas above a solution con-
taining less than 3% ethanol.
6.
Press or click the start button to begin recording data.
The probe stabilizes after about 10 minutes of warm up time.
Ethanol Sensor
1
2
®
PS-2194
Introduction
The Ethanol Sensor measures the concentration of ethanol in a
gas or indirectly in a liquid. The sensor's probe contains a mate-
rial whose resistance decreases in the presence of ethanol. The
sensor translates this resistance into an ethanol concentration
measurement, which is recorded and displayed by the connected
computer or datalogger. The Ethanol Sensor is well-suited for
continuous recording and discrete measurements.
Precautions
Do not expose the probe to gas containing more than 3%
ethanol, including the gas above alcoholic beverages.
Higher ethanol concentrations will deplete the sensing ele-
ment. Note that the gas above a liquid contains about the
same concentration of ethanol as the liquid itself.
Ensure that the membrane is installed and intact before
using the probe. The membrane prevents the sensing element
from getting wet. If the sensing element accidentally gets
wet, rinse and dry it quickly. Note that the probe may be
degraded or destroyed.
Do not immerse the probe. The membrane (if it is in good
condition) will protect the probe in case of accidental immer-
sion. However, the probe is designed for use in gas only.
Connecting the Parts
1.
Plug the amplifier's PASPORT connector into any port of a
PASPORT interface or datalogger.
2.
Connect the probe to the amplifier.
Connecting the probe and interface
3.
If you are using a computer, connect the PASPORT interface
to it and start DataStudio.
I n s tr u c t i o n S he e t
probe
0 1 2 - 1 0 2 8 7B
to PASPORT
interface

Advertisement

loading

Summary of Contents for Pasport PS-2194

  • Page 1 However, the probe is designed for use in gas only. Quick Start Connecting the Parts Connect the amplifier to your PASPORT interface. Plug the amplifier’s PASPORT connector into any port of a PASPORT interface or datalogger. Connect the probe to the amplifier. Connect the probe to the amplifier.
  • Page 2 The probe contains a heating element. After connecting the sen- Change the membrane when it is torn or discolored. sor to a powered PASPORT interface, wait about 10 minutes for the sensor’s temperature to stabilize. Pull the retaining ring off the end of the probe.
  • Page 3 PASCO sci- entific, is prohibited. Trademarks PASCO, PASCO scientific, and PASPORT are trademarks or registered trademarks of PASCO scientific, in the United States and/or in other countries. All other brands, products, or service names are or may be trademarks or service marks of, and are used to identify, products or services of, their respective owners.