Pasport PS-2190 Instruction Sheet

Pasport PS-2190 Instruction Sheet

1 mhz 2-channel voltage sensor

Advertisement

Quick Links

1 MHz 2-channel Voltage Sensor
Included Equipment
1 MHz 2-channel Voltage Sensor
Additional Equipment Required
Xplorer GLX datalogger (PS-2002)
Patch Cords (SE-9750 or SE-9751)
Quick Start
1.
Connect the 1 MHz 2-channel Voltage Sensor to your
PASPORT Xplorer GLX datalogger.
2.
Use patch cords to connect Channel 1 and Channel 2 to the
parts of the circuit for which you want to measure the volt-
age.
3.
On the Xplorer GLX, press the start button (s) to begin
recording data.
4.
When data recording is done, press the start button again to
stop recording data.
The sensor measures up to two voltages simultaneously at speeds
up to one million samples per second when used with the GLX in
oscilloscope mode.
®
PS-2190
Introduction
The 1 MHz 2-channel Voltage Sensor is a high speed, two chan-
nel PASPORT voltage sensor that has a basic voltage range of
±10 volts. It has amplifiers that allow full-scale input ranges of
±1 volt and ±0.1 volt. The sensor uses an oversampling scheme
that allows input sampling at speeds up to one million samples
per second.
The 1 MHz 2-channel Voltage Sensor connects to a PASPORT
Xplorer GLX datalogger. Measurements are recorded and dis-
played by the datalogger.
Set-up
Connecting the Sensor to an Xplorer GLX
Datalogger
Plug the sensor into any port on the top of the PASPORT Xplorer
GLX datalogger. The Xplorer GLX screen will automatically
show a Graph display of voltage versus time.
Setting Sensor Parameters
Sample rate, data averaging, and input voltage range are set in the
Sensors screen.
Press the Home button (h) to open the Home screen.
Press F4 (I) under Sensors in the Home screen to open
the Sensors screen.
Instruction Sheet
Figure 1: Sensor screen
012-10253A

Advertisement

loading

Summary of Contents for Pasport PS-2190

  • Page 1 Connecting the Sensor to an Xplorer GLX • Patch Cords (SE-9750 or SE-9751) Datalogger Plug the sensor into any port on the top of the PASPORT Xplorer GLX datalogger. The Xplorer GLX screen will automatically Quick Start show a Graph display of voltage versus time.
  • Page 2 1 M H z 2 - c h a n n e l V o lt a g e S e n s o r PS-2190 The Sensors screen shows the icon and name of the sensor in the Selecting Scope Mode...
  • Page 3 1 M H z 2 - c h a n n e l V o l t a g e S e n s o r P S - 2 1 9 0 The sample rate is linked to the time resolution on the horizontal Use the up arrow cursor button to increase the Trigger Level.
  • Page 4 The PS-2190 has two BNC connectors on the rear panel labeled Trigger In and Trigger Out. By connecting the Trigger Out of one PS-2190 to the Trigger In of a second PS-2190, the two sensors are time synchronized and four channels can be measured. Since the Xplorer GLX can currently only display two channels in Scope Mode, this feature is for future use.