Mod. V895 16 Channel Leading Edge Discriminator

3.7. Test pulse

(Base address + %4C write only)
A test pulse on all output channels can be generated by performing a write access at
Base address + %4C; the test pulse is generated independently from the number written
into this register.

3.8. Module identifier words

(Base address + %FA, + %FC, + %FE, read only)
Three words located at the Base address + %FA,+ %FC, + %FE of the page are used to
identify the module, as shown in Fig. 3.1:
15
V e r s i o n
The word located at the address Base + %FE identifies the single module via a serial
number, and any change in the hardware will be shown by the version number .
For the Mod. V895 the word at the address Base + %FC has the following configuration:
Manufacturer N°
Type of module
NPO:
00101/97:V895x.MUTx/03
14
13
12
11
10
9
M o d u l e ' s
Manufacturer number
% F A
F i x e d
c o d e

Fig. 3.1: Module Identifier Words

=
000010 b
=
0001010100
Filename:
v895_rev3.doc
Revision date:
19/03/09
8
7
6
5
4
3
s e r i a l
n u m b e r
M o d u l e
t y p e
% F 5
F i x e d
c o d e
Revision:
3
2
1
0
Address
Base + % FE
Base + % FC
Base + % FA
Number of pages:
25
Page:
19