External Connectors; Input Connectors; Output Connectors - Caen V895 Series Manual

16 channel leading edge discriminators
Table of Contents

Advertisement

Document type:
Title:
User's Manual (MUT)
Mod. V895 16 Channel Leading Edge Discriminator

2.4. External connectors

The location of the connectors is shown in Fig. 2.1. Their function and electromechanical
specifications are listed in the following subsections.

2.4.1. INPUT connectors

INPUT CHANNELS:
VETO INPUT:
TEST INPUT:

2.4.2. OUTPUT connectors

OUTPUT CHANNELS:
OR OUTPUT:
NPO:
00101/97:V895x.MUTx/03
Filename:
v895_rev3.doc
Revision date:
19/03/09
Mechanical specifications:
16 LEMO 00 type connectors.
Electrical specifications:
negative polarity, 50 Ohm impedance, DC
coupling; input range: -5 mV ÷ -5 V; input
offset: ±5 mV; 140 MHz maximum input
frequency.
Mechanical specifications:
1 LEMO 00 type connectors.
Electrical specifications:
standard NIM logic signal, high impedance,
15 ns minimum FWHM; leading edge of the
VETO signal must precede of at least 8 ns the
leading edge of the input and overlap
completely the input signal; the VETO signal
doesn't act on TEST input.
Mechanical specifications:
1 LEMO 00 type connectors.
Electrical specifications:
standard NIM logic signal, high impedance,
5 ns minimum FWHM, 60 MHz maximum
input frequency.
Mechanical specifications:
4 Header 3M 3408-D202 type, 8+8 pin
connectors.
Electrical specifications:
Differential
ECL
level
impedance; pulse width adjustment from 5±1
ns to 40±5 ns FWHM.
Input/Output delay: 15.5±1.5 ns.
Mechanical specifications:
1 LEMO 00 type connectors.
Electrical specifications:
standard NIM logic signal, 50 Ω impedance;
50 MHz maximum input frequency.
Revision:
3
on
110
Ohm
Number of pages:
Page:
25
12