Modules
Multiplexer 34925A
Terminal Block 34925T
The overcurrent protection devices will be rated, or the snubber circuits will limit the current, according to:
Overcurrent Protection Device Rating per Application Voltage (1.6VA max) = 0.02A
1. Peak voltage, channel-to-channel or channel-to-earth.
2. DC or peak AC current.
3. Limited to 6 W of channel resistance power loss per module.
4. With input protection resistors: 2 x 100
range.
5. Into analog bus. System errors are included in the internal DMM measurement accuracy specifications.
6. Channel resistance is typically < 1.5
current. Increased relay channel resistance for measurements with load currents below 10 mA can occur on cards that have been out of
service or following relay inactivity for periods of greater than 1 week. Switching relays for 2K cycles prior to use may reduce the variation in
channel resistance. Applies to the 34921A and 34922A. Keysight recommends the use of 4-wire Ohms for resistance measurements. For
high accuracy voltage measurements, select the DMM input resistance setting of >10 G ohms to minimize the impact of relay contact
resistance.
90
Pollution Degree 1
40/80 channels, 1MHz
1
±80Vpeak
2
0.02A
3
1.6VA/channel
7
Volt-Hertz limit: 10
Initial closed channel resistance
<700
±5%; 0.5W; TC = ±200 ppm/°C. The series resistance of the 34925A limits the use of the 100
but can go as high as 50
Pollution Degree 2
40/80 channels, 1MHz
1
±80Vpeak
2
0.02A
1.6VA/channel
Volt-Hertz limit: 10
4,5,6
:
Initial closed channel resistance
<700
when a channel is used in measurement applications with < 10 mA load
Keysight 34921A-34925A User's Guide
3
7
4,5,6
:
Transients
1000Vpk
Need help?
Do you have a question about the 34922A and is the answer not in the manual?