▶ Do not replace detachable mains supply cords For maintenance, refer to Sparklike Oy or with inadequately rated cords. Sparklike Service Partners ▶ Always make sure that you are using a safety...
For connecting or disconnecting the cables, refer to sparklike.com/guidance/lp. Connector panel: Connections 3. STARTING UP THE SPARKLIKE LASER After connecting all the cables and turning the The power switch has an indicator light: The power switch has an indicator light: main switch on, turn on the device from the power switch on the main unit.
4.2 Software functionality and pages Double glass measurement is set as a default for Sparklike Laser analyser, this can be observed The following chapters will discuss each page of the from the “Double IG” indicator being green. It is software –...
Page 4
In the middle of the page, you will also be able values via the drop-down boxes, the software identi- to observe the results on the thickness of the fies the glass and cavities thicknesses in the ±1 mm glass and the space between the glasses as range from each given value.
This ▶ Figures on the graph should be equally ensures the correct operation of Sparklike Laser. spaced and between 30-70 points spacing To ensure the quality of the measurements it is approximately. See test menu picture.
Page 6
(light on the button will turn on). This will open the internal valves and guide the flushing gas Sparklike Laser is a gas fill analyser; however, it flow through the measuring head. During the measures the oxygen absorption from the insulating filling, press “Self check”...
Page 7
Safe values to edit are, e.g. High_limit and Low_limit for the Measuring head indicator lights: Measuring head indicator lights: indicator lights on the Main page of Sparklike Application. It is not recommended to edit Purple light continuous...
Page 8
Main page with sample measurement results for triple glass the vacuum by pressing the measuring head 5.5 Factors to consider when measuring against the glass firmly once the measurement has started. If this does not help, firstly clean Here is a list of factors that might impact the the vacuum with a dry cloth.
DAQ card is working. Then shut down windows and after that, restart the device. If this does not solve the problem, contact the product support at Sparklike for instructions. If DAQ led is off, contact the product support at Sparklike for instructions.
Page 10
“Laser I&R” test, one can fine-tune the temperature correct, the device is firmly against the glass, and as follows (Sparklike technical support can help you the device is not tilted. In these cases, the device on this).
Page 11
(otherwise keep the measuring head vertically on the holder) Sparklike Laser is a gas fill analyser that works by measuring the oxygen inside the glazing unit. ▶ If all of these are OK, but the problem...
7. SERVICE AND MAINTENANCE 7.1 Calibration To ensure the measurement accuracy Sparklike Laser Portable 2.1 must be calibrated and mainte- nanced annually. To request calibration and/or maintenance, please contact Sparklike Service Center sparklike.com/en/care-and-maintenance/ calibration. 7.2 Technical support using remote connection Sparklike can offer remote technical support using TeamViewer application.
4x circular multipin connector, RJ-45, 2x USB 2.0, VAC Power (C14) Display 12” HD Touch Screen Software Windows 10, Laser Software by Sparklike (all included) Data logging capability SSD Storage (15 GB), USB/Network transfer * Typical construction of the insulating glass unit...
Page 14
Scan the QR-code for further instructions and support SPARKLIKE OY www.sparklike.com Helsinki, Finland 05/2023...