3.2 Connections of VP Dual coax cable WU10-V-D, WF10 and WE10
pH conventional is where combined sensor with high ohmic pH membrane and low ohmic
WU10-V-D,
WF10, WE10
Coax1 core
Coax1 shield
Coax2 core
Coax2 shield
Wire
Wire
Wire
Wire
Drain wire
WU10-V-D,
WF10, WE10
Coax1 core
Coax1 shield
Coax2 core
Coax2 shield
Wire
Wire
Wire
Wire
Drain wire
Parameter ORP (pH compensated, for pH conventional)
WU10-V-D, WE10
Coax1 core
Coax1 shield
Coax2 core
Coax2 shield
Wire
Wire
Wire
Wire
Drain wire
Parameter pH conventional + ORP
internal reference.
Color
VP pin
brown
A
brown
B
white
C
white
D
red
E
blue
F
yellow
G
green
H
black
shield
Parameter ORP (for pH conventional)
Color
VP pin
brown
A
brown
B
white
C
D
white
red
E
blue
F
yellow
G
green
H
black
shield
Color
VP pin
brown
A
brown
B
white
C
D
white
red
E
blue
F
yellow
G
green
H
black
shield
EXA/FLXA terminal no.
15
13
16
14
11
12
63 (4W analyzer only)
EXA/FLXA terminal no.
13
15 & 14
(interconnected)
11
12
63 (4W analyzer only)
EXA/FLXA terminal
no.
13
17
15 & 14
(interconnected)
11
12
63 (4W analyzer only)
Description
pH input
Ref input
pH guard
Liquid Earth (LE)
Pt1000-T1
Pt1000-T2
Not connected
Not connected
Shield
Description
Not connected
Ref input
Not connected
Metal / Liquid
Earth (LE)
Pt1000-T1
Pt1000-T2
Not connected
Not connected
Shield
Description
pH input
Ref input
pH guard
Liquid Earth (LE)
Pt1000-T1
Pt1000-T2
Not connected
Not connected
Shield
IM 12B06W02-02EN-P
13