Service Department of the heavy industrial environment. relevant Yokogawa Sales office for repair or replacement (at Yokogawa’s discretion). This multicore shielded cable is available in The following information must be included...
1.4 Serial number The Serial number is defined by nine (9) alphanumeric characters: Production location Year/Month code Tracking number Example: N3Y295015 Table 1: Production Year code Year Year Year code Year code 2014 2026 2015 2027 2016 2028 2017 2029 2018 2030 2019...
2. GENERAL SPECIFICATIONS 2.1 WU10 Single coax cable Model WU10-V-S-..-.. Type Single coax cable with VarioPin (VP) connector and wire pin finish Construction Multicore shielded cable with one low noise coax and three insulated wires and drain wire Tabel 3: WU10-V-S single coax cable properties, part 1 Model WU10-V-S-..-..
Page 6
Tabel 4: WU10-V-S single coax cable properties, part 2 Model WU10-V-S-..-.. Description: Single coax cable Electrical specifications: Resistance: (@ 20°C) ≤ 40 Ω/km Lead 19 x 0.18mm ≤ 80 Ω/km Lead 7 x 0.2 mm ≤ 45 Ω/km Drain wire Capacitance: •...
2.2 WU10/WF10 Dual coax cable Model WU10-V-D-..-.. WF10-… Type Dual coax cable with Cable without connector (-FF) for use Vario pin and wire pin finish as extension cable with BA10 terminal extension box Construction Multicore shielded cable with two low noise coaxes and four insulated wires Tabel 5: WU10-V-D and WF10 dual coax cable properties, part 1 Model...
Page 8
Tabel 6: WU10-V-D dual coax cable properties, part 2 Model WU10-V-D-..-.. and WF10-… Type: Dual coax cable Electrical specifications: Resistance: (@ 20°C) ≤ 45Ω/km Lead 19 x 0.14mm ≤ 45Ω/km Drain wire Capacitance: • Coax (conductor-shield) ≤ 180 pF/m • Coax (conductor-overall shield) ≤...
2.3 WE10 Dual coax cable XL Model WE10-H-D-...-.. Type Dual coax cable XL Female connector with wire pins for direct connection to terminals With female-male connector as extension cable or without connectors for use as extension cable with BA10 terminal extension box Construction Multicore shielded cable with two low noise coaxes and four insulated wires, halogen free.
2.4 Environmental conditions Storage temperature : -10°C to +50 ºC (14 to 122 ºF) Ingress Protection VP connector : IP67 (conform IEC 60529) Hazardous areas : The WU10/WF10 and WE10 cables can be used in an explosive atmosphere and hazardous areas when used with appropriate accessories and certified sensors.
3. INSTALLATION 3.1 Connections of VP Single coax cable WU10-V-S Parameter pH Conventional pH Conventional is a combined sensor with high ohmic pH membrane and low ohmic internal reference. For ORP and pH-differential measurement the WU10-V-S cannot be used. For these measurements the WU10-V-D or WE10 is necessary WU10-V-S Color VP pin...
3.2 Connections of VP Dual coax cable WU10-V-D, WF10 and WE10 Parameter pH conventional + ORP pH conventional is where combined sensor with high ohmic pH membrane and low ohmic internal reference. WU10-V-D, Color VP pin EXA/FLXA terminal no. Description WF10, WE10 Coax1 core brown...
Page 14
Parameter pH differential + ORP pH differential is where the combined sensor has two dissimilar glass membranes WU10-V-D, Color VP pin EXA/FLXA terminal no. Description WF10, WE10 Coax1 core brown pH input Coax1 shield brown pH guard Coax2 core white Ref input Coax2 shield white...
Page 15
Parameter SC WU10-V-D, Color VP pin EXA/FLXA terminal no. Description WF10 Coax1 core brown Ui (inner electrode) Coax1 shield brown Ii (Inner electrode) Coax2 core white Uo (Outer electrode) Coax2 shield white Io (Outer electrode) Wire Pt1000-T1 Wire blue Pt1000-T2 Wire yellow Not connected...
Page 16
Parameter DO - galvanic WU10-V-D, Color VP pin EXA/FLXA terminal no. Description WF10 Coax1 core brown Cathode Coax1 shield brown Anode Coax2 core white Not connected Coax2 shield white Not connected Wire Pt1000-T1 Wire blue Pt1000-T2 Wire yellow Membrane check Wire green Not connected...
5. MODEL CODES Model Code Suffix code Description WU10 Sensor cable Connection type Variopin Single coax Cable type Dual coax 2 meters 5 meters Cable length 10 meters 15 meters 20 meters Model Code Suffix code Description WF10 Connecting cable -005 up to and including Length in meter between 005 Length in m...