12

3. INSTALLATION

3.1 Connections of VP Single coax cable WU10-V-S

pH Conventional is a combined sensor with high ohmic pH membrane and low ohmic internal
reference. For ORP and pH-differential measurement the WU10-V-S cannot be used.
For these measurements the WU10-V-D or WE10 is necessary
WU10-V-S
Color
Coax core
brown
Coax shield
brown
Wire
black
Wire
red
Wire
blue
Drain wire
yellow
For pH differential the WU10-V-S cannot be used, for this parameter the WU10-V-D or WE10 is
For SC the WU10-V-S cannot be used, for this parameter the WU10-V-D or WE10 is
For ISC the WU10-V-S cannot be used, for this parameter the WU10-V-D or WE10 is
WU10-V-S
Color
Coax core
brown
Coax shield
brown
Wire
black
Wire
red
Wire
blue
Drain wire
yellow
WU10-V-S
Color
Coax core
brown
Coax shield
brown
Wire
black
Wire
red
Wire
blue
Drain wire
yellow
For LDO the WU10-V-S cannot be used, for this parameter the WU10-V-D or WE10 is
IM 12B06W02-02EN-P
Parameter pH Conventional
VP pin
EXA/FLXA terminal no.
A
B
C
E
F
D
Parameter pH differential
necessary
Parameter SC
necessary
Parameter ISC
necessary
Parameter DO-galvanic
VP pin
EXA/FLXA terminal no.
A
B
C
E
F
D
Parameter DO-Polarographic
VP pin
EXA/FLXA terminal no.
A
B
C
E
F
D
Parameter LDO
necessary
Description
15
pH input
13
Ref input
16
pH guard
11
Pt1000-T1
12
Pt1000-T2
14
Liquid Earth (LE)
Description
13
Cathode
15
Anode
16
Membrane check
11
Pt1000-T1
12
Pt1000-T2
14
Shield
Description
17
Cathode
18
Anode
11
Pt1000-T1
12
Pt1000-T2
14
Shield/Liquid Earth (LE)