INF RATE: Enter the infusion rate, press YES
9.
10. UP Ramp: Enter the up ramp time (if any), press YES
11. DN Ramp: Enter the down ramp time (if any), press YES
12. TOT TIME: Enter the total time for the infusion, press YES. Continue
pressing YES if needed until the KVO field is highlighted.
13. KVO RATE: Enter KVO (Keep Vein Open) rate, then press YES (0 to 10 ml/hr)
14. DONE?: Press YES to accept
When at the Run to Start Screen, press the HELP/OPTIONS key to set the Lock Level
(security level), and to change the Air-In-Line sensitivity and the Down Occlusion
sensitivity from the default settings when the therapy requires these changes:
Lock: Press YES to accept or NO to change Lock setting, press YES to accept new setting.
DN Occlu: Press YES to accept (default setting is LOW) or
press NO to change then press YES to accept.
AIR SENS: Press YES to accept (default setting is 0.5 mL)
or NO to change then press YES to accept.
Press YES at each field to advance to the next field. Press
YES to ACCEPT OPT at the end of the menu.
PROGRAMMED BY:
CHECKED BY:
tel: 801.264.1001
toll free: 800.970.2337
www.curlinpump.com

Inf RATE:

UP Ramp:

DN Ramp:

TOT TIME:

KVO RATE:

Lock

DN Occlu:

AIR SENS:

DATE:
DATE:
CURLIN® is a registered trademark of Curlin Medical, Inc. Moog® is a
registered trademark of Moog, Inc. © 2018 Moog, Inc. Moog Medical, 4314
Zevex Park Lane, Salt Lake City, UT 84123 USA. LS-61476 Rev. A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CURLIN 6000 and is the answer not in the manual?

Questions and answers

Table of Contents