Technical specifications
8.1 General technical specifications
Measuring accuracy
•
•
Rated conditions
Ambient temperature
•
•
Storage temperature
Degree of protection (according to EN 60529)
Electromagnetic compatibility
Mechanical construction
Weight
Process connections
Electrical connections
Material of the wetted parts
•
•
•
Non-wetted parts material
•
26
Influence of vibration (per IEC 60068-2-6)
Effect of auxiliary power supply
Height
Relative humidity
Measuring cell
Process connection
Sealing material
Viton (FPM)
Neoprene (CR)
Perbunan (NBR)
EPDM
Enclosure
0.005%/g to 500 Hz in all directions
0.005%/V
-25 ... +85 °C (-13 ... +185 °F)
Max. 2 000 m above mean sea level
Use a suitable power supply at an altitude of more
than 2 000 m above mean sea level.
0 ... 100%
‑50 ... +100 °C (-58 ... +212 °F)
•
IP65 with plug according to EN 175301-803-A
•
IP67 with M12 plug
•
IP67 with cable
•
IP67 with fast-fit cable gland
•
In accordance with EN 61326-1
•
In accordance with EN 61326-2-3
•
In accordance with NAMUR NE21, only for ATEX
devices and with a maximum measurement de‐
viation of ≤1%
Approx. 0.090 kg (0.198 lb)
Dimension drawings (Page 29)
•
Plug in accordance with EN 175301-803-A shape A with cable
entry M16x1.5 or ½-14NPT or Pg 11
•
M12 plug
•
2- or 3-wire (0.5 mm
) cable (∅ 5.4 mm)
2
•
Fast-fit cable gland
Stainless steel, mat. no. 1.4435
Stainless steel, Mat. No. 1.4404 (SST 316 L)
15th digit in the order number
A
B
C
D
Stainless steel, Mat. No. 1.4404 (SST 316 L)
Compact Operating Instructions, 09/2020, A5E03304388-AA
Medium temperature
-15 ... +125 °C (+5 ... +257° F)
-35 ... +100 °C < 100 bar (‑31 ...
+212 °F; < 1 450 psi)
-20 ... +100 °C (‑4 ... +212 F)
-40 ... +125 °C < 100 bar (‑40 ...
+257 °F; < 1 450 psi), can be used
for drinking water
SITRANS P210