Bedfont ToxCO Quick Start Manual
Hide thumbs Also See for ToxCO:

Advertisement

Quick Links

Infection Control and
Maintenance Guidelines
For Bedfont® ToxCO® Monitor
Before using any of the above equipment, the Operating Manual should be read and fully
understood.
Infection Control
Washing hands before and after testing is highly recommended for both operator and user as
part of a sensible infection control regime.
The monitor uses a SteriBreath™ mouthpiece that connects to the monitor via a D-piece™.
SteriBreath™ mouthpieces are single-use only as re-use can increase the risk of cross-infection.
The D-piece™ contains a one-way valve to prevent patients drawing air back from the monitor.
An integrated infection control filter removes and traps >99% of airborne bacteria and >97% of
viruses
.
1
Bedfont® recommends that the D-piece™ is replaced every 28 days or if visibly soiled. The
D-piece™ cannot be sterilised as this process may affect the positioning of the valve. It is
possible that saliva can build up inside the D-piece™ if they are not replaced regularly, which
may over time affect the valve and efficacy of the filter material.
It is preferable that, where possible, the user attaches their own SteriBreath™ mouthpiece to
the D-piece™ before the breath test, and detaches and disposes of it once the test is complete.
Whilst the user is exhaling, the operator should avoid positioning themselves in front of the
exhaust port of the instrument.
Routine maintenance
See the operating manual for full maintenance instructions.
The D-piece™ requires changing at least every 28 days or if visibly soiled. A reminder will
appear on the ToxCO® start-up screen.
Batteries should be replaced when indicated. They should be removed if the instrument is not
likely to be used for some time.
Bedfont® Scientific Ltd.
Station Road, Harrietsham, Maidstone,
Kent, ME17 1JA England
Tel: +44 (0)1622 851122 Fax: +44 (0)1622 854860
Email: ask@bedfont.com Web: www.bedfont.com
© Bedfont® Scientific Limited 2022
Issue 4 - August 2022, Part No: LAB286_USA_TOXCO
Bedfont® Scientific Limited reserves the right to change or update his literature without prior notice.
Registered in: England and Wales. Registered No: 1289798
Stephen Rowe
Cristimar E4-1
Ave Juan Carlos I
Los Cristianos, Arona, 38650
Santa Cruz de Tenerife, Spain
Rx Only

Advertisement

loading
Need help?

Need help?

Do you have a question about the ToxCO and is the answer not in the manual?

Questions and answers

Summary of Contents for Bedfont ToxCO

  • Page 1 An integrated infection control filter removes and traps >99% of airborne bacteria and >97% of viruses Bedfont® recommends that the D-piece™ is replaced every 28 days or if visibly soiled. The D-piece™ cannot be sterilised as this process may affect the positioning of the valve. It is possible that saliva can build up inside the D-piece™...
  • Page 2 Maintenance Guidelines Calibration • The ToxCO® should be calibrated with 50ppm CO in air; an onscreen reminder will display every 6 months, however the ToxCO® should be calibrated at least once a year • It is recommended that only Bedfont® calibration gas is used; the unit may give false readings if not calibrated.