MGE UPS Systems SAM Operating And Maintenance Manual page 9

Portable suction

Advertisement

When using the SAM e.p.s unit via the ac supply mains it is important -
The location of the SAM e.p.s must always allow for the removal of the power lead to enable the unit to be
isolated from the mains power supply if necessary and it must be located on a firm and stable base avoiding
causing any trip hazard.
Operation of the SAM e.p.s
Having charged the battery fully, prepared the unit with the desired collection vessel and filter, the unit is
ready for use. The unit may be operated from its internal battery or from an external mains supply. The SAM
e.p.s will work without a battery when connected to a supply mains. The SAM e.p.s will always require a reset
after the battery is disconnected or reconnected - this is done by turning the unit ON and OFF.
Connect the patient tubing to the 'PATIENT' inlet of the collection vessel.
Turn the unit ON by setting the control knob to the desired vacuum level.
When the suction operation is completed set the control knob to 'Off'.
Operating temperature - 5 C to +45 C
After Care of the SAM e.p.s
The collection vessel should be emptied when it reaches 750ml.
Disconnect the filter connection from the collection vessel.
Remove the collection vessel from the cradle.
Remove the patient tubing from the collection vessel.
Safely dispose of the contents of the collection vessel according to local protocols.
Clean the SAM e.p.s according to the instructions ref: maintenance section
It is recommended that the filter capsule be changed after each days use, or immediately if it becomes
wetted by froth, or, after aspiration of any infective material.
Prepare the SAM e.p.s for its next patient.
Carrying the SAM e.p.s
When carrying the SAM e.p.s using the shoulder strap, be careful not to snare the unit on any
restrictions or obstructions along the pathway.