Guidance and manufacturer's declaration – electromagnetic immunity
The SAM e.p.s is intended for use in the electromagnetic environment specified below. The customer or the user of the
SAM e.p.s should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC61000-4-2
Electrical fast transient / burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage
dips,
interruptions
and
variations on power supply
input lines
IEC61000-4-11
Power frequency (50Hz)
Magnetic field
IEC61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Manual 156 - Issue 08
SAM e.p.s – UK
Operation and Maintenance Manual
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
short
<5% U
T
voltage
(>95 % dip in U
For 0.5 cycle
40% U
T
(60 % dip in U
T)
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5% U
T
(>95 % dip in U
For 5 s
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
Input / output line tests
not applicable.
± 1 kV differential mode
± 2 kV common mode
<5% U
T
(>95 % dip in U
T)
T)
For 0.5 cycle
40% U
T
(60 % dip in U
T)
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5% U
T
)
(>95 % dip in U
)
T
T
For 5 s
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the SAM e.p.s requires continued
operation during power mains
interruptions, it is recommended
that the SAM e.p.s be powered
from an uninterruptible power
supply or a battery.
If incorrect operation occurs, it
may be necessary to position the
SAM e.p.s further from sources of
power frequency magnetic fields
or to install magnetic shielding.
The power frequency magnetic
field should be measured in the
intended installation location to
assure that it is sufficiently low.