4.3 MULTI-CHANNEL PIPETTE RANGE
8 Channels & 12 Channels
Color
CAT No.
Code
EASD-G10
EASD-T10
EASD-G20
EASD-T20
EASD-G50
EASD-T50
EASD-10G
EASD-10T
EASD-20G
EASD-20T
EASD-30G
EASD-30T
4.4 SETTING THE VOLUME
Delivery volume is clearly indicated in the
volume display seen in the main body of
the pipe e.
In variable volume models, the bo om
volume wheel includes a small increment
scale for precise setpoint and delivery
capabili es.
Some variable volume pipe es include one
or two decimal places in the setpoint volume
wheels. This is indicated by the use of a black
horizontal line as seen in the following
examples sec on.
Incre-
Volume
ments
Range
[ μl ]
[ μl ]
0.5-10 μl
0.02
2-20 μl
0.02
5-50 μl
0.1
10-100 μl
0.1
20-200 μl
0.2
30-300 μl
0.2
- 5 -
Test
Inac-
Volume
curacy
[ μl ]
(±) %
1
8
5
4
10
2
2
7
10
3
20
2
5
3
25
1.5
50
1
10
3
50
1
100
0.8
20
5
100
1
200
0.7
30
3
150
1
300
0.6
Impreci-
sion
(±) %
5
2
1
3
2
1.6
2
1
0.7
2
0.8
0.3
1.4
0.4
0.25
1
0.5
0.3
1
0
0
0