Do not use the 8714D with the model 8712H or 8782 transmitters. Damage to the calibrator will result. NOTICE This document provides basic guidelines for the Rosemount 8714D. It does not provide instructions for detailed configuration, diagnostics, maintenance, service, troubleshooting. This QSG is also available electronically on Emerson.com.
Page 4
Emerson Flow customer service Note If NIST traceability is required, then Emerson recommends the 8714D be factory calibrated on a yearly basis. If you need to return your device for calibration, or if the device does not operate properly, then contact Flow Support at one of the email links shown below.
September 2022 Quick Start Guide Step 1: Change transmitter parameters 1. Use a Field Communicator or Local Operator Interface (LOI) to change the parameters of the transmitter to the following: Tube Calibration Number: 1000015010000000 Units: ft/s Analog Output range: 20 mA = 30.00 ft/s Analog Output Zero: 4 ma = 0 ft/s Coil Pulse Mode:...
Wall mount Use the cable assembly with the six contact pin flying leads. Emerson part number 08714-0205-0003 Field mount Use the cable assembly with the two black plastic amp connectors. Emerson part number 08714-0205-0002 1. Insert the metal end of the wiring assembly into the 8714D Calibration Standard.
Page 7
September 2022 Quick Start Guide Wall mount transmitters Use the six contact pin terminal plug-style wiring assembly. Follow the numeric convention for the 8712 so the plugs match the terminal block. Figure 3-1: 8712C/8712D/8712ES Terminal block Figure 3-2: 8712EM/8750WDMW Terminal blocks 9 10 19 18 11 12...
Page 8
Quick Start Guide September 2022 Field mount transmitters If you do not have a LOI, the coil and electrode connectors will be clearly visible on the bottom side of the board. If you have a LOI, remove the three mounting screws from the LOI assembly and pull the assembly away from the circuit board beneath.
Rosemount 8714D. 1. Set the Rosemount 8714D to simulate a flow rate of 30 ft/s. 2. Power up the transmitter with the Rosemount 8714D connected. Allow the electronics 30 minutes to warm-up before reading the flow rate.
Page 10
Quick Start Guide September 2022 ™ Rosemount 8714D (Calibration Standard) Magnetic Flowtube Simulator...
Page 11
September 2022 Quick Start Guide Quick Start Guide...