Contact Information ........... 4 Checklist before calling/e-mailing for support..4 Shipping ............. 4 1.3 The calibration kit includes ........5 2 The BVM CAL kit ............. 6 2.1 BVM CAL Box ............. 6 Battery power ............. 6 Inserting batteries ..........6 2.2 Other parts in the kit ...........
1.2 Service and support This manual is valid for the BVM Cal Box kit and it is For technical assistance please contact your local designed for use for calibration of the Megger BVM Megger representative or direct your request to the instruments.
BVM calib sw Keysight (CJ-8110X) Control cable, black, RJ45, 0.5 m GA-01100 Extension cable, black, 0.5 m 04-30050 Test cable, red, 0.5 m 04-35012 Test cable, black, 0.5 m 04-35010 User’s manual CJ0125XE Transport case GD-00620 CJ0125BE ZP-CJ01E BVM CAL KIT...
2 THE BVM CAL KIT The BVM CAL kit 2.1 BVM CAL Box Insert the batteries, observe polarity. Input from Power and Signal Connector (To 1st BVM unit) (RJ45 cable, Order No. GA-01100) Output to BVM unit (IN) unit for calibration (RJ45 cable, Order No.
2 THE BVM CAL KIT 2.2 Other parts in the kit 2.3 Other equipment needed BMV CAL Ref Digital Multimeter (DMM) The system is designed for use together with Agilent 34401A (connected via serial / COM-port) or Agilent 34410A (LAN / Ethernet connected). The system can also be used with Keysight 34461A (LAN / Ethernet connected).
Power & Signal Connector V/Ω Digital Multimeter BVM CAL Ref Agilent 34401A Unit for Power COM-port calibration supply unit To mains Figure 3.2.1 Connection diagram for calibration of BVM units using instrument with COM port BVM CAL KIT ZP-CJ01E CJ0125BE...
Power & Signal Connector V/Ω Digital Multimeter BVM CAL Ref Agilent 34410A Unit for Power calibration supply Keysight 34461A unit Ethernet-port To mains Figure 3.2.2 Connection diagram for calibration of BVM units using instrument with Ethernet port CJ0125BE ZP-CJ01E BVM CAL KIT...
You can try to run the calibration once again to see if you get the same result. Click “Reject” and then “New Unit” to run a new calibration. A defective unit should be replaced. Click “Yes”. The calibration will start. BVM CAL KIT ZP-CJ01E CJ0125BE...
If the battery power is to weak you will get the fol- Bad connection lowing error message. Here are some examples showing error messages when a connecetion is lost. Check batteries and replace with new batter- ies. Check all leads and connections. CJ0125BE ZP-CJ01E BVM CAL KIT...
Connect in the same way as for a standard calibration but shift positions for CAL Ref BVM and BVM unit. Run a normal calibration. Shift back to right position for the CAL Ref. BVM CAL KIT ZP-CJ01E CJ0125BE...
Page 13
4 SPECIFICATIONS CJ0125BE ZP-CJ01E BVM CAL KIT...
AC adapter 115 X 52 X 30 mm Transport case 270 X 180 X 240 mm Weight BVM Cal Box 0.5 kg BVM CAL Kit 2.1 kg (Incl. transport case) Output Output voltage 20 V DC (max) BVM CAL KIT...
Page 15
4 SPECIFICATIONS CJ0125BE ZP-CJ01E BVM CAL KIT...
Page 16
▪ Insulation Power Factor (C&DF) Test Equipment worldwide. ▪ Insulation Resistance Test Equipment Megger is certified according to ISO 9001 and 14001. ▪ Line Testing Equipment Megger is a registered trademark. ▪ Low Resistance Ohmmeters ▪...
Need help?
Do you have a question about the BVM Cal kit and is the answer not in the manual?
Questions and answers