Esd Guidelines - HP 85097A User Manual

Electronic calibration system
Table of Contents

Advertisement

Operating an ECal System
Electrostatic Discharge (ESD)

ESD Guidelines

Use the following guidelines before connecting devices, modules, or test
ports.
• Always have a grounded antistatic mat in front of your test
equipment and wear a grounded wrist strap attached to it (see
Figure
2-1).
• Ground yourself before you clean, inspect, or make a connection to a
static-sensitive device or test port. You can, for example, grasp the
grounded outer shell of the VNA port connector briefly to discharge
static from your body.
• Discharge static electricity from a device before connecting it: touch
the device briefly (through a resistor of at least 1 MΩ) to either the
outer shell of the test port, or to another exposed ground. This
discharges static electricity and protects test equipment circuitry.
Refer to the Electronic Calibration Module Reference Guide (HP part
number 85091-90008) for the HP part numbers of ESD supplies.