LEAD IN WATER TEST PROCEDURE
1
PREPARE SAMPLE
Add 50mL of water sample to the 50mL conical tube.
Then, add fi ve (5) drops of eXact Reagent ACID-1, mix,
and allow to sit for at least 5 minutes.
2
TURN METER ON
Press the
LEADQuick
3
SELECT TEST
Press and re-press the
display shows HG.
4
FILL CELL
Fill and empty the cell 4 times with the TEST
SAMPLE prepared in Step 1. Finally,
capacity to begin test. Tip meter to allow excess
sample to fl ow out in order to make room for
Pb-2 Reagent addition.
5
ADD REAGENT
Add fi ve (5) drops of eXact Reagent Pb-2 to
CELL.