READY SNAP
1. Turn on the eXact
three times with clean water, and empty CELL.
2. Select the PB MENU.
3. Snap open one Ready Snap
Standard by twisting the top 180°. NOTE: DO NOT CON-
SUME THE STANDARD.
4. Squeeze plastic ampoule to fi ll photometer CELL to
capacity (4mL).
5. Discard this liquid sample followed with a quick shake of
the meter to empty the CELL of the remaining water drops.
6. Squeeze plastic ampoule to fi ll photometer CELL to
capacity (4mL).
7. Run the test method as you do normally using the correct
procedure (pages 4-5).
8. Compare the displayed value against the Assigned Value
chart provided below.
9. If your value is within the Acceptable Range, then you are
operating the eXact
10. If your value is Borderline, review the Instruction Manual
and verify that you are following the proper procedure.
Repeat the test again from Step 1 through Step 8.
Menu
Parameter
Test
PB
Lead in
Water
TEST PROCEDURE
®
LEADQuick Photometer, rinse CELL
®
®
LEADQuick Photometer correctly.
®
Desired
Value
130
ppm
leadquickmeter.com
1P Method Verifi cation
Acceptable
Range
5
Borderline
Range