10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Guidance and manufacturer's declaration – electromagnetic immunity
The model MIT-W102 SERIES is intended for use in the electromagnetic environment specified below. The customer
or the user of the model MIT-W102 SERIES should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3,8
12
IEC
Compliance
60601
level
test
level
Portable and mobile RF communications equipment
should be used no closer to any part of the model
MIT-W102 SERIES, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
3
d = 1,2
Vrms
150
Vrms
kHz
to 80
d = 1,2
MHz
d = 2,3
V/m
3 V/m
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
80
manufacturer and d is the recommended separation
MHz
distance in meters (m).
to 2,5
GHz
45
3,8
12
Electromagnetic environment – guidance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
7,3
23