5-3-3. Liquid sterilization

Applicable chemicals
General name
Hydrogen peroxide
Sterilization procedure
Washing
Soaking in
chemical solution
Washing with
water
Warning
After chemical sterilization, thoroughly wash the probe with sterilized water.
Residual chemicals can cause an adverse reaction on the bodies of the operator or patient.
Trade name
TM *
PERASAFE
Practical liquid 1.62W/V%
Be sure to wash and fully dry the probe before sterilization.
See section 5-1.
Soak the insertion portion and handle in a chemical solution at the
shortest time where effective sterilization can be achieved.
Sufficiently rinse the probe using sterilized water to remove
chemicals.
-35-
* Except Canada