Sterilization; Ethylene Oxide Gas (Eog) Sterilization - Hitachi UST-678 Instruction Manual

Transrectum composite probe
Table of Contents

Advertisement

5-3. Sterilization

Perform Ethylene oxide gas (EOG) sterilization, STERRAD sterilization or Liquid sterilization as necessary.

5-3-1. Ethylene oxide gas (EOG) sterilization

Applicable gases
General name
Ethylene oxide gas
Sterilization procedure
Washing
EOG sterilization
Aeration
Warning
Perform full aeration after gas sterilization.
Residual gas can cause an adverse reaction on the bodies of the operator or patient.
Trade name
TM
AMPROLENE
84% density
Be sure to wash and fully dry the probe before sterilization.
See section 5-1.
Sterilize the following conditions:
Temperature:
Humidity:
Compression
Decompression
Aerate the following conditions to ensure that no gases remain.
Temperature:
Humidity:
Compression
Decompression
Period:
-33-
MN1-5255 Rev. 15
Manufacturer
Central Uni Co., LTD.
10℃ to 60℃
50°F to 140°F
30% to 85%
to +101.175kPa
to -100.000kPa
10°C to 60°C
50°F to 140°F
30% to 85%
to +101.175kPa
to -100.000kPa
5 days (natural aeration)