Sterrad Sterilization - Hitachi UST-678 Instruction Manual

Transrectum composite probe
Table of Contents

Advertisement

MN1-5255 Rev. 15
5-3-2. STERRAD
Applicable gases
General name
Hydrogen peroxide
(58% density)
Sterilization procedure
Washing
STERRAD
sterilization
Remarks
Some discoloration of the probe may occur, but this does not affect performance or safety.
Caution
Do not sterilize a probe without a STERRAD label* using the STERRAD system.
It can cause damage or deterioration to the probe shell. Use the STERRAD system only to sterilize
a probe with a STERRAD label.
Do not put the probe directly into the sterilization pouch*.
This can cause the film of the sterilization pouch to stick to the cable covering, resulting in damage
to the cable. Completely wrap the entire probe (including the transducer, cable and connector) with
sterilization wraps* before putting it into the sterilization pouch
*: A Johnson & Johnson company Division of Ethicon, Inc product
®
sterilization
Trade name
STERRAD
Sterilization system
Be sure to wash and fully dry the probe before sterilization.
See section 5-1.
ADVANCED STERILIZATION PRODUCTS
A Johnson & Johnson company
* STERRAD label
-34-