Hitachi UST-678 Instruction Manual page 27

Transrectum composite probe
Table of Contents

Advertisement

① Connect one end of the extension tube to the inlet/drain nipple of the probe, and connect the other end of
the tube to the three-way cocks. Either of the three-way cocks may be connected.
② Insert the probe into the balloon.
③ Put the proper amount of deaerated water in the syringe, then
connect the syringe to the extension tube.
④ To evacuate air from inside of the probe, inject the proper amount
of deaerated water from the syringe to the balloon.
⑤ Remove bubbles in the tip of the balloon.
⑥ Attach a rubber band supplied as an accessory, on the groove of the
probe (Part A of the figure).
⑦ Remove excessive water from the top of the balloon.
⑧ If bubbles still remain in the balloon, inject water from the syringe
into the balloon. Using the same method as that used for draining
water from the inlet/drain port of the probe (Part B of the figure),
remove bubbles from the balloon.
[Remarks]
Surgical tape is included in the accessory pack. When the flares of
the balloon disturb to use, fasten the flare by this tape.
Syringe
-21-
MN1-5255 Rev. 15
Balloon
Inlet/drain port nipple
Tube
rubber band
A