Gas Disinfection - Hitachi UST-678 Instruction Manual

Transrectum composite probe
Table of Contents

Advertisement

MN1-5255 Rev. 15

5-2-2. Gas disinfection

Applicable gases
General name
Formalin gas
Disinfection procedure
Washing
Gas disinfection
Aeration
Warning
Perform full aeration after gas disinfection.
Residual gas can cause an adverse reaction on the bodies of the operator or patient.
Trade name
F. gen
(14% formaldehyde)
Be sure to wash and fully dry the probe before disinfection.
See section 5-1.
Disinfection the following conditions:
Temperature:
Humidity:
Atmospheric pressure:
Aerate the following conditions to ensure that no gases remain.
Temperature:
Humidity:
Compression
Decompression
Period:
-32-
Manufacturer
Aso Pharmaceutical Co., Ltd.
Room temperature
Normal humidity
Normal pressure
10°C to 60°C
50°F to 140°F
30% to 85%
to +101.175kPa
to -100.000kPa
5 days (natural aeration)