Transportation and installation
(5-4) device installation
Get the device out of its box carefully. To do so, take out
1
all the protective material (Plastofoams) around it, take out
the steam cleaner, and put it on a flat and standard
laboratory table such that the space within 25 cm radius of
the device is empty. Make sure that the surface and legs of
the table are not made of metal.
If the surface or legs of the table is metal and is in contact
with Earth, There will be a risk of electric shock. Therefore,
place the Poli VAC on the wooden or fiberglass.
Polivac should not be exposed to sunlight directly.
2
An appropriate cord should be used to connect the device
3
to its nearby power supply.
An electrical protector must be put on the way of the
4
socket by the user, so overload or indirect currents don't
have contact with the protective parts of the device.
Installation of brush holder on the motor shaft
Put one of the brush holder ends, which has a hole, on the
motor shaft, and push it inward slowly.
Push the other end of the brush holder inward with a plastic
hammer until it is fixed properly.
Rotate the motor shaft and brush holder manually, and make
sure it is installed correctly and brush holder axis is in parallel
with the shaft. Also, make sure that the motor shaft and brush
holder axes are aligned with each other.
In order to check the installation of the device being correct,
turn it on, and if any deviation or problems of this kind are
seen, try to fix it. If the deviation is more than 2-1 mm, open
that part and install it again.